BM1VOM-1.24
7. Temperature 89
BM1VET Operation Manual
ALARM LIMIT
Setting numeric value is 15.0
℃
~ 45.0
℃
.
1. Move the
▶
mark to select either RETURN or TEMP, and press.
2. After pressing the cursor at TEMP, move it to LOW, and press.
3. When the color has changed, move the cursor again to select a target value, and press.
4. Move the cursor to HIGH and press. After the color has changed, move the cursor again
to select a target value, and press. (One may choose HIGH first to get the same result.)
5. Select RETURN to get out of the menu.
ALARM
LIMIT
PREV
MENU
ALARM
SOUND :
ON
MAIN
MENU
TEMPERATURE ALARM LIMIT
RETURN
UNITS
LOW
HIGH
TEMP
°C
30.0
42.0