Rev. 5.01
87
ECG Cable color
AHA
:
American Heart Association (U.S.A.)
IEC
:
International Electrotechnical Commission (Europe)
3LEAD / 5LEAD
Lead wire
AHA
Color code
AHA
Label
IEC
Color code
IEC
Label
Right arm
White
RA
Red
R
Left arm
Black
LA
Yellow
L
Right leg
Green
RL
Black
N
Left leg
Red
LL
Green
F
V1(precordial)
Brown
V1
White
C1
ECG Signal Processing and Display
The monitor displays QRS Complex for an amplitude of 0.4 to 5.0 mV (0.2 -5.0 mV for scale
settings below 0.5 mV / cm) and a QRS width of 70 to 120 ms for large animals (also for small
animals with 40 to 100 ms). The heart rate is calculated from 15 to 300 times per minute using
the last 10 seconds of the R-R interval and the two longest intervals and the two shortest
intervals at the R-R interval. The remaining interval is averaged, and the current heart rate is
displayed in the HR parameter box of the main screen as a result.
If arrhythmia monitoring is possible, the HR parameter box will change accordingly. If you select
Basic, you can display three basic life treatening arrhythmias called ASYS, VFIB, and VTAC. If Full is
selected, a separate ARR parameter box will be displayed next to the HR parameter box (for
details on selecting the arrhythmia mode, refer to the Arrhythmia Setting chapter).
Содержание BM Elite Series
Страница 21: ...Rev 5 01 21 Left side view Printer ...
Страница 25: ...Rev 5 01 25 Left side view Printer ...
Страница 29: ...Rev 5 01 29 Left side view Printer USB port ...
Страница 40: ...Rev 5 01 40 BM5Vet Elite BM7Vet Elite 1 2 3 4 1 2 3 4 ...
Страница 57: ...Rev 5 01 57 Display Mode Wave and numeric mode Numeric Mode ...
Страница 76: ...Rev 5 01 76 Pop up ST Window BM5 7Vet Elite BM3Vet Elite ...
Страница 84: ...Rev 5 01 84 Connector and Measurement cable ECG connector BM3Vet Elite BM5Vet Elite BM7Vet Elite ...
Страница 86: ...Rev 5 01 86 ECG Lead 5 Positions of 5 Lead Placement RL V LL LA RA 3 Positions of 3 Lead Placement LL LA RA ...
Страница 93: ...Rev 5 01 93 top waveform position 3 Select Pacemaker 4 SELECT PACEMAKER ON ...
Страница 102: ...Rev 5 01 102 Connector and Measurement Cable SpO2 connector BM5 7Vet Elite BM3Vet Elite ...
Страница 110: ...Rev 5 01 110 BM5Vet Elite BM7Vet Elite Cuff for USA only ...
Страница 160: ...Rev 5 01 160 EtCO2 malfunction Failure in battery recharge In case the battery will not charge in 6 hours or more ...
Страница 161: ...Rev 5 01 161 Power failure Print failure ...
Страница 162: ...Rev 5 01 162 Dual Gas Malfunction Data storage failure ...
Страница 163: ...Rev 5 01 163 Periodic Noises ...
Страница 215: ...Rev 5 01 215 BIONET CO LTD ...