5. Alarm
Overview
The monitor displays the alarm limits (parameter threshold) and can be configured by the user to
raise an alarm if exceeded. Limits are displayed both in the alarm limits table and in the parameter
box. If this limit is exceeded, a visual and/or audible alarm will occur.
The bedside monitor is the primary alarm device, and there may be other secondary alarm devices
depending on how you configured the device / network. Depending on the alarm condition, the
monitor generates an alarm using one or more of the following devices:
⚫
sound reflecting alarm severity
⚫
Change the color in the parameter box of the alarm parameter
⚫
Alarm messages in the local message area
⚫
Alarm banner indicating alarm status
⚫
External alarm device such as nurse call system
⚫
Activate alarm recording
The monitor generates an alarm when the parameter in the Alarm Limits table is
ON
. It is not a
prerequisite that the parameter is displayed on the display or connected in the event of an alarm.
Alarm Priority
The alarm type is divided into an animal status alarm and a product status alarm.
The animal status alarm sounds when the diagnostic function (ECG 13 auto diagnosis) and alarm
upper and lower limits are exceeded, and there are levels of High, Medium, Low and Message, and
there is a difference in the order and volume of the alarm. You can set the alarm level for each
parameter and function. The animal status alarm provides the highest priority alarm.
The features of each alarm are described as follows. The alarm priority is High > Medium > Low >
Message. For high or medium alarms, the printer output is supported when
‘Ala
rm print on
’
is set.
Содержание BM Elite Series
Страница 21: ...Rev 5 01 21 Left side view Printer ...
Страница 25: ...Rev 5 01 25 Left side view Printer ...
Страница 29: ...Rev 5 01 29 Left side view Printer USB port ...
Страница 40: ...Rev 5 01 40 BM5Vet Elite BM7Vet Elite 1 2 3 4 1 2 3 4 ...
Страница 57: ...Rev 5 01 57 Display Mode Wave and numeric mode Numeric Mode ...
Страница 76: ...Rev 5 01 76 Pop up ST Window BM5 7Vet Elite BM3Vet Elite ...
Страница 84: ...Rev 5 01 84 Connector and Measurement cable ECG connector BM3Vet Elite BM5Vet Elite BM7Vet Elite ...
Страница 86: ...Rev 5 01 86 ECG Lead 5 Positions of 5 Lead Placement RL V LL LA RA 3 Positions of 3 Lead Placement LL LA RA ...
Страница 93: ...Rev 5 01 93 top waveform position 3 Select Pacemaker 4 SELECT PACEMAKER ON ...
Страница 102: ...Rev 5 01 102 Connector and Measurement Cable SpO2 connector BM5 7Vet Elite BM3Vet Elite ...
Страница 110: ...Rev 5 01 110 BM5Vet Elite BM7Vet Elite Cuff for USA only ...
Страница 160: ...Rev 5 01 160 EtCO2 malfunction Failure in battery recharge In case the battery will not charge in 6 hours or more ...
Страница 161: ...Rev 5 01 161 Power failure Print failure ...
Страница 162: ...Rev 5 01 162 Dual Gas Malfunction Data storage failure ...
Страница 163: ...Rev 5 01 163 Periodic Noises ...
Страница 215: ...Rev 5 01 215 BIONET CO LTD ...