is 15~20seconds.
C-1.
Module reset
EtCO2 module initializing.
Note
For best result, connect the CAPNOSTAT 5 CO2 Sensor to an adapter and wait
2 minutes before performing the Adapter Zero procedure.
Status Message
Following is a list of some of the message that may appear on the monitor when monitoring CO2.
The message should clear when normal operating criteria are met or a solution is found.
Status Message
Cause
Solution
SENSOR OVER
TEMP
The sensor temperature is greater
than 40
°
C
Make sure sensor is not exposed
to extreme heat(heat lamp,etc.)
SENSOR FAULTY
One of the following conditions
exists:
Capnostat Source Current Failure
EEPROM Checksum Faulty
Hardware Error
Check that the sensor is properly
plugged in. Reinsert or reset the
sensor if necessary.
SENSOR WARM
UP
Sensor under temperature,
Temperature not stable,
Source Current unstable
This error condition is normal at
startup. This error should clear
when the warm up is complete.
CHECK SAMPLING
LINE
This error occurs whenever the
pneumatic pressure is outside the
expected range.
Check that the sampling line is not
occluded or kinked. Replace the
sample line
ZERO REQUIRED
Zero Required, Zero Error
To clear, check airway adapter and
clean if necessary. If this does not
correct the error, perform an
adapter zero. If you must adapter
zero more than once, a possible
Содержание BM Elite Series
Страница 21: ...Rev 5 01 21 Left side view Printer ...
Страница 25: ...Rev 5 01 25 Left side view Printer ...
Страница 29: ...Rev 5 01 29 Left side view Printer USB port ...
Страница 40: ...Rev 5 01 40 BM5Vet Elite BM7Vet Elite 1 2 3 4 1 2 3 4 ...
Страница 57: ...Rev 5 01 57 Display Mode Wave and numeric mode Numeric Mode ...
Страница 76: ...Rev 5 01 76 Pop up ST Window BM5 7Vet Elite BM3Vet Elite ...
Страница 84: ...Rev 5 01 84 Connector and Measurement cable ECG connector BM3Vet Elite BM5Vet Elite BM7Vet Elite ...
Страница 86: ...Rev 5 01 86 ECG Lead 5 Positions of 5 Lead Placement RL V LL LA RA 3 Positions of 3 Lead Placement LL LA RA ...
Страница 93: ...Rev 5 01 93 top waveform position 3 Select Pacemaker 4 SELECT PACEMAKER ON ...
Страница 102: ...Rev 5 01 102 Connector and Measurement Cable SpO2 connector BM5 7Vet Elite BM3Vet Elite ...
Страница 110: ...Rev 5 01 110 BM5Vet Elite BM7Vet Elite Cuff for USA only ...
Страница 160: ...Rev 5 01 160 EtCO2 malfunction Failure in battery recharge In case the battery will not charge in 6 hours or more ...
Страница 161: ...Rev 5 01 161 Power failure Print failure ...
Страница 162: ...Rev 5 01 162 Dual Gas Malfunction Data storage failure ...
Страница 163: ...Rev 5 01 163 Periodic Noises ...
Страница 215: ...Rev 5 01 215 BIONET CO LTD ...