Bioness Ness H200 Скачать руководство пользователя страница 1

Clinician's Reference Card

Fitting the H200 Wireless Orthosis

Rx Only

 

(US Only)

1

3

2

5

4

6

7

8

9

10

Select the appropriate Orthosis size and then attach the appropriate Thenar.
Attach the appropriate Wrist Insert.
Don the Orthosis and check between the FPL #5 electrode and the wrist. If there is a gap, 
attach and tighten the Orthosis Wrist Strap. 
If there is still a gap, doff the Orthosis and attach the FPL Panel.
Select and attach an Extensor Fitting Panel that will elicit the desired hand/finger movement. 
Start with fitting panel A.
Select and attach a Flexor Fitting Panel that will elicit the desired hand/finger movement. 
Start with fitting panel A.
Wet all of the cloth electrodes.
Attach the cloth electrodes to the electrode bases.
Don the Orthosis.
Adjust stimulation parameters.

612-00373-001 Rev. D
11/2015

Adjusting Stimulation Parameters

1

3

2

Log into the H200 Wireless Software. Select a patient record and press Open, or press New 
to create a patient record.
From the Programs Menu, press Stim. Parameters.
Adjust stimulation parameters for the extensors and flexors. Press Start to turn on 
stimulation. Press Stop to stop stimulation. Deactivate the trigger button on the Orthosis, if 
appropriate. Press Done to return to the Programs menu.

3

2

1

Deactivate Remote 

Trigger

1

Thenar

4

FPL Panel

5

Extensor Fitting Panel

7

9

8

10

6

Flexor Fitting 
Panel

3

Wrist Strap

Wrist Insert

2

For a complete list of all contraindications, warnings, adverse reactions, precautions, and care instructions please review the 

Clinician’s Guide found at www.bioness.com.