7

Step 2:

Preparing the System to Begin Treatment 

Gently press electrodes onto the skin. Depending on your ability to 
move after your surgery, it may be helpful to ask another person 
to help you place these electrodes on your skin. See page 14 for 
electrode instructions. Consult your local Biomet representative 
should you have questions about your electrode placement. Make 
sure you apply the electrodes after your skin has been cleaned and 
dried. If you skin becomes abnormally red at the electrode sites, 
the electrodes should be moved adjacent to the original sites. 
If redness does not go away after 48 hours with the electrodes 
moved, you should contact your prescribing physician.

Place the stimulator 
controller in a 
comfortable and 
convenient location for 
you to wear for your 
prescribed treatment 
period. If you decide 
to place the stimulator 

controller in the holster, place the holster in a comfortable and 
convenient location that is close to your fracture site.

Connect the two electrode wires into the electrode lead wire. Insert 
the electrode lead wire plug into the jack at the top of the stimulator
controller (Figure 1). Choose the electrode lead wire that reaches 
from the location of the device or device holster to your fracture 
site (where your electrodes should be placed). The device holster 
can be worn on a belt, attached to your waistband, or clipped 

onto the extremity band.

Remove the fully charged 
battery pack from the 
battery charger cradle by a 
gentle lift (2) of the small 
tab (1). The battery is now 
ready for connection to 
the Biomet

®

 OrthoPak

®

 

Non-invasive Bone Growth 
Stimulator System. 

Insert the battery pack into 
the Biomet

®

 OrthoPak

®

 

Non-invasive Bone Growth 
Stimulator System by 
placing it into the battery 
compartment (1) (replace the 
cover by sliding back to its 
closing position). The cover 

     

 

will snap into place (2).

2

3

1

Содержание OrthoPak

Страница 1: ...Biomet OrthoPak Non invasive Bone Growth Stimulator System Patient Manual ...

Страница 2: ...his procedure bone is taken from another location of the patient s body and surgically implanted at the nonunion site Metal rods or plates and bone screws also may be attached to the broken bone to secure the ends in place while it mends An alternative nonsurgical procedure was developed for the treatment of nonunions This method incorporated the transmission of an electrical signal through the no...

Страница 3: ...ns Wearing the Biomet OrthoPak Non invasive Bone Growth Stimulator System The Biomet OrthoPak Non invasive Bone Growth Stimulator System has been designed so that it is convenient to use comfortable to wear and safe to operate You should begin using the Biomet OrthoPak Non invasive Bone Growth Stimulator System device immediately after you have read the directions for use Directions for Use It is ...

Страница 4: ...pe Electrode Covers Covers are to be used for non casted applications by patients who are experiencing difficulty keeping the electrode pads in good contact with the skin When the electrode retainer is properly applied over the electrode pad you will be able to shower without removing the electrode pad Figure 2 Device Holster The device holster is designed to securely hold the Biomet OrthoPak Non ...

Страница 5: ...vice it is recommended that you take the second battery pack place it into the charger and charge fully In the meantime you may use the first battery pack to begin your treatment immediately NOTE The first battery pack may not provide a 24 hour treatment initially Additional batteries are available through Biomet customer Service Each day preferably at the same time you should change the battery f...

Страница 6: ...s unit in any other device Use of the Biomet OrthoPak Non invasive Bone Growth Stimulator System battery packs in any other device may cause damage or malfunction to the batteries and or devices Do not short circuit overcharge crush mutilate nail penetrate heat reverse the or terminals or disassemble the battery pack Do not allow metal objects to come into contact with the battery pack terminals T...

Страница 7: ...er is appropriately plugged into the A C outlet Remove the current battery pack from the Biomet OrthoPak Non invasive Bone Growth Stimulator System and place that battery pack in the charger cradle for charging Open the battery compartment by pressing the battery cover release button 1 and sliding the battery cover to open 2 4 5 Step 1 Continued Battery Charging ...

Страница 8: ...cide to place the stimulator controller in the holster place the holster in a comfortable and convenient location that is close to your fracture site Connect the two electrode wires into the electrode lead wire Insert the electrode lead wire plug into the jack at the top of the stimulator controller Figure 1 Choose the electrode lead wire that reaches from the location of the device or device hols...

Страница 9: ... your sales representative Helpful Tips Loose Electrodes Make sure that both electrodes are in complete contact with your skin Moisten or replace worn electrodes if necessary Incomplete Circuit Check all connection points insuring tight fit of lead wire plug into the Biomet OrthoPak Non invasive Bone Growth Stimulator System device and full engagement of electrode cable pin into electrode wire rec...

Страница 10: ...ck is already fully charged the charger will terminate the recharging process early This will be indicated by the charger having no orange light when charging is complete Do not be concerned if the battery pack is kept in the charger cradle for a long period of time When the Biomet OrthoPak Non invasive Bone Growth Stimulator System Needs a Freshly Charged Battery Pack the Following will Occur 1 T...

Страница 11: ...r will not treat until USB cable is disconnected Contact Biomet Troubleshooting Electrodes Change your electrodes as required Different skin types will provide for a longer or shorter life of the electrodes If the alarm indicates a disconnection it is likely that either the lead wire connection is incomplete or the adhesive gel on the electrode is no longer working and the electrodes need changing...

Страница 12: ...cords the number of days you receive treatment This helps your doctor track your treatment Helpful Tips Keep the audible alarm setting in the ON position as much as possible This alarm will help warn you of any problems with the device During special occasions when you would like the device not to tell you audibly about stimulator problems you may press the button for 3 seconds to turn off the aud...

Страница 13: ...rly If You Have Questions If you have questions about your Biomet OrthoPak Non invasive Bone Growth Stimulator System contact BIOMET domestically at 1 800 526 2579 or 1 973 299 9300 if calling from outside the United States Representatives are available from 8 30am to 5 00pm EST Monday through Friday At other times please leave a clear message for a return call by the next business day IMPORTANT A...

Страница 14: ...allows full weight bearing on the casted extremity unless gross motion greater than 5 degrees in any plane at the nonunion site is present In such case weight bearing is not advised and should not be permitted as this may compromise the effectiveness of the treatment The safety and effectiveness of the use of the device on individuals lacking skeletal maturity has not been established In the prese...

Страница 15: ...ction generally resolves spontaneously following diagnosis and correction of the underlying cause Electrode Instructions for Use Do not open outer packet until ready to use 1 Tear open packet 2 Remove electrode from clear plastic backing liner 3 Wet finger with tap water and moisten entire gel area 4 Place electrode on skin 5 Connect electrode to electrode lead Renewal 1 With continuous use electr...

Страница 16: ...ith EBI lineage Biomet has helped over one million people To learn more about this product contact your local Biomet Sales Representative today PN 1067800 00 Rev D Biomet OrthoPak Non invasive Bone Growth Stimulator System Patient Manual 399 Jefferson Road Parsippany NJ 07054 800 526 2579 www biomet com BNS231001L 03 14 2014 EBI LLC All trademarks are the property of Biomet Inc or one of its subsi...

Отзывы: