10
Disconnection
System Error
End of operation/
Treatment Completion
Low Battery Charge
Symbol Condition
Instruction
Treating
Continue use.
Audible alarm
engaged
If beeping, depress the
button briefly to silence the
alarm. Depress the button
approximately 3 seconds to
engage or disengage the
audible alarm.
Insert a charged battery pack.
Confirm that each electrode is
properly applied on the skin.
See Page 14 for instructions.
Confirm that the lead wire is
attached properly. Replace the
lead wire if necessary.
Error in the stimulator –
Contact Biomet for assistance.
Stimulator is
connected to a
PC
Stimulator will not treat until
USB cable is disconnected.
Contact Biomet
Troubleshooting - Electrodes
• Change your electrodes as required. Different skin types will
provide for a longer or shorter life of the electrodes. If the alarm
indicates a disconnection, it is likely that either the lead wire
connection is incomplete or the adhesive (gel) on the electrode
is no longer working and the electrodes need changing.
Check all lead wire connection points, to make sure that the
electrode lead wire is tightly plugged into the top of the Biomet
®
OrthoPak
®
Non-invasive Bone Growth Stimulator System (See
figure below) and that the lead wire connectors are completely
inserted into both electrode connectors. If all the connections
are made and the symbol indicates a disconnection, it is
probably time to change the electrodes.