BIOMET Clotalyst Скачать руководство пользователя страница 1

Artist Proof: 01-15-10

Clotalyst

®

Autologous Serum  

Collection System

Содержание Clotalyst

Страница 1: ...Artist Proof 01 15 10 Clotalyst Autologous Serum Collection System ...

Страница 2: ...raft or allograft for bone graft handling prior to application to an orthopedic surgical site The Clotalyst Autologous Serum Collection System produces approximately 6ml of autologous serum from only 12ml of patients own citrated blood The serum output can be mixed with platelet rich plasma PRP produced by the GPS III system ...

Страница 3: ...sary to produce PRP The Clotalyst System is designed to clot PRP in less than 15 seconds The kit produces 6ml of PRP and 6ml of Clotalyst autologous serum Platelet gel derived from the GPS III System in combination with Clotalyst Autologous Serum Clotalyst Autologous Serum ...

Страница 4: ...ent syringe from sterile pouch and attach plunger Note Leave grey stopper on opposite end of reagent syringe when attaching plunger Remove grey stopper on reagent syringe Unscrew blue cap on port No 2 of disposable and attach reagent syringe While holding the Clotalyst disposable vertically inject all of reagent Remove the reagent syringe from the disposable Remove protective cover on white cap an...

Страница 5: ...Unscrew red cap on port No 3 and attach syringe containing 12ml of citrated blood While holding disposable vertically inject all blood Remove the empty syringe Remove protective cover on the white cap and discard Screw white cap onto port No 3 ...

Страница 6: ...ter The glass beads inside the disposable should be distributed evenly when it is placed into the heater Press the incubation button to start cycle When the incubation cycle is complete 25 minutes the heater will beep four times Step 4 Spin Preparation At the end of the 25 minute incubation period shake the Clotalyst disposable to dissolve any gel present Technique continued ...

Страница 7: ...nd time to 5 minutes Press green button to start centrifuge cycle Once cycle is complete press red button to open lid Step 6 Autologous Serum Extraction Gently remove the disposable from the centrifuge without disturbing the stratified layers Unscrew the yellow cap from the center port while holding the disposable vertically Connect sterile 12ml syringe to the center port and slowly extract 5 to 6...

Страница 8: ... No 3 as autologous serum is extracted Step 7 Storage Store Clotalyst autologous serum at 18 26 C 64 78ºF for up to 4 hours prior to use Discard remaining serum if not used within the provided time frames Step 6 Autologous Serum Extraction continued Technique continued ...

Страница 9: ... TIP Accessories Description Catalog Number Clotalyst Kit includes GPS III Separator with Blood Draw and Reagent Clotalyst Kit includes GPS III Mini Separator with Blood Draw and Reagent 800 0722 800 0723 Clotalyst Heater Heater Cord included 800 0755 Clotalyst Non Sterile Counterbalance Not included in Clotalyst kit 800 0760 Drucker 230 Volt 50 60Hz Centrifuge 755VES Clotalyst System Ordering Inf...

Страница 10: ...nt and possible adverse effects 7 The autologous serum exhibits thrombin activity and therefore must not be injected into or otherwise allowed to enter the patient s vascular system 8 The safety and effectiveness of the device for bone healing and hemostasis has not been established 9 Follow manufacturers instructions when using a Biomet Biologics centrifuge IEC centrifuge or the Drucker company c...

Страница 11: ...talyst Separator during steps 3 and 4 can occur This is NOT detrimental to the procedure 4 LOAD BLOOD ENSURE BLOOD FROM ONLY ONE PATIENT IS PROCESSED Unscrew and discard red cap on port 3 Attach and slowly transfer contents of anticoagulated blood filled syringe 1ml ACD A mixed with 11ml of patient s whole blood through port 3 Unscrew and discard protective inner piece from white cap tethered to p...

Страница 12: ...dicated This material is intended for the sole use and benefit of the Biomet Biologics sales force and physicians It is not to be redistributed duplicated or disclosed without the express written consent of Biomet For product information including surgical techniques indications contraindica tions warnings precautions and potential adverse effects see the Biomet Biologics website at www biometbiol...