8/19/2010 

 

MICROCON®  

 

EXC7 and EXC7-UV 

 

 

User Manual 

 

Read and Save these Instructions

 

 

 

 
 
 
 
 
 
 
 
 
 

Manufactured by 

 

 
749 Hope Road, Suite A- Eatontown, NJ 07724 USA   
Tel: 800-224-9768, 732-389-8922   Fax: 732-389-8821 
 www.biologicalcontrols.com 

 

Содержание MICROCON EXC7

Страница 1: ...0 MICROCON EXC7 and EXC7 UV User Manual Read and Save these Instructions Manufactured by 749 Hope Road Suite A Eatontown NJ 07724 USA Tel 800 224 9768 732 389 8922 Fax 732 389 8821 www biologicalcontr...

Страница 2: ...Copyright 2010 Biological Controls 8 19 2010...

Страница 3: ...N THIS MANUAL AND ATTACHED TO THE UNIT BEFORE STARTING INSTALLATION OR MAINTENANCE ACTIVITIES IMPROPER INSTALLATION ADJUSTMENT ALTERATION SERVICE MAINTENANCE OR USE CAN CAUSE FIRE ELECTRICAL SHOCK OR...

Страница 4: ...Copyright 2010 Biological Controls 8 19 2010 Blower Module Filter Module Control Panel Interconnect Cable Assembly Filter Module...

Страница 5: ...Plate Control Unit optional 4 3 2 5 Key Switch optional 5 4 SYSTEM START UP 5 5 MAINTENANCE INFORMATION 6 5 1 Blower Module 6 5 1 1 Blower Removal 6 5 2 Filter Module 6 5 2 1 Cleaning 7 5 2 2 Hour Me...

Страница 6: ...contains a 1 5hp hp motor or hp blower motor All of the configurations are capable of producing either negative or positive room pressure environments The system is typically equipped with a single Bl...

Страница 7: ...of duct types and sizes is also an option 2 2 Blower Module Mechanical The Blower Module can be mounted on a platform or suspended Typically it is mounted above a suspending ceiling grid and supporte...

Страница 8: ...the air more evenly throughout a room 3 INSTALLATION ELECTRICAL Turn off power at the facility fuse or circuit breaker panel before proceeding with wiring or installation activities All electrical in...

Страница 9: ...4VDC power module is hard wired to be on 24 7 and there is no power off switch The Blower Module High Voltage must be turned off at the facility circuit breaker panel or the Interconnect Cable Assembl...

Страница 10: ...to gain access When the system components have been mounted and all of the duct work and electrical wiring connections have been completed the system is ready for operation Use the on off switch loca...

Страница 11: ...metal screws that connect the blower exhaust port to the blower chassis Remove the six 6 20 hex head bolts holding the blower motor mounts to the chassis The blower along with the motor mounting feet...

Страница 12: ...ule chassis The meter does not have a reset function to prevent inadvertent resets Always log the meter readings on the Filter Module label after maintenance activity 5 3 Filter Renewal 5 3 1 Pre Filt...

Страница 13: ...o the Filter Module Check the HEPA gasket to make sure it is compressible and not damaged The filter media is very fragile and can be damaged easily Be careful not to touch or puncture the HEPA filter...

Страница 14: ...nes call for pressure differential of negative 010 H2 O W C or minus 2 49 Pascal s The room must be sealed well and leak free especially the ceiling to create a negative pressure environment This allo...

Страница 15: ...period the product is defective or malfunctions Biological Controls or its dealer or distributor from whom the product was purchased shall at the option of Biological Controls replace or repair the d...

Страница 16: ...Copyright 2010 Biological Controls 8 19 2010 11 Manufacturer TEL 800 224 9768 BIOLOGICAL CONTROLS FAX 732 389 8821 749 Hope Road Suite A WEB SITE www biologicalcontrols com Eatontown NJ 07724...

Страница 17: ...Copyright 2010 Biological Controls 8 19 2010 12 9 SPECIFICATIONS 9 1 Filter Module Feature Description 9 2 Blower Modules...

Страница 18: ...Copyright 2010 Biological Controls 8 19 2010 13 10 ASSEMBLY FIGURES DIMENSIONS 10 1 Filter Module...

Страница 19: ...Copyright 2010 Biological Controls 8 19 2010 14 10 2 Blower Module...

Страница 20: ...Copyright 2010 Biological Controls 8 19 2010 15 11 SCHEMATICS 11 1 Typical System 115V Schematic 11 2 Typical System 230V Schematic...

Страница 21: ...Copyright 2010 Biological Controls 8 19 2010 16 11 3 Typical System 1 5 HP 2 Speed Motor Schematic 11 4 Remote Wall Switch Option Schematics...