Operating the BioSpot BT600 

6.1

 

Start up 

The operation software, including all required drivers is pre-installed on the device! 
The license information for the Windows 7 OS can be found on the backside of the device. 
See the license sticker! 

Before operation of the BioSpot BT600 workstation using the BioFluidix software, please read the 
additional “

BioFluidiX Control Software BETA Manual V1.0” 

carefully! 

 

1.

 

Connect a mouse and keyboard to the BioSpot BT600 backside USB connectors. Connect a 
screen to the backside DVI connector. 

2.

 

Connect the BioSpot BT600 to a standard power supply socket (90 - 240V). 

o

 

Fuse >= 6A, B- C-characteristic 

3.

 

Make sure the emergency switch is unlocked  

4.

 

Switch the backside power toggle switch to “1” (on) to start the BioSpot BT600 

o

 

the internal PC will start up automatically 

o

 

switch on the screen, Windows 7 (OS) will start 

o

 

login with the BFX profile (no password required) 

5.

 

Start the BioFluidix Software (BFX) clicking on the link on the desktop  

o

 

the control software to operate the BioSpot BT600 will start 

o

 

the axes start automatically to reference towards the lower left corner 

6.2

 

Connection of PipeJets, reservoirs and aspiration unit 

 

Use a 2 mm allen key to move the reservoir 

holder upwards  

 

Cut a 3-4cm long piece of Tygon LMT-55 and 

mount it on the fluidic connector of the 

PipeJet pipe and connect it to the 300 µl 

intermediate reservoir.  

3-4cm Tygon tubing 

 

 

 

  

 

1ml intermediate reservoir 

Содержание BioSpot BT600

Страница 1: ...BT BioFluidix GmbH Engesserstra e 4a D 79108 Freiburg Germany www biofluidix com User Manual BioSpot BT600 Version 1 1 Date March 2018...

Страница 2: ...ist of parts 5 5 Description of the BioSpot BT600 6 5 1 BioSpot BT600 workstation 6 5 2 Connectors at the control electronics located at the z axisFehler Textmarke nicht definiert 5 3 Backside connect...

Страница 3: ...es on cables or damages on mechanical parts If it is supposed that it is not possible to launch a safe operation you have to take it out of service and protect it against an unintended operation It is...

Страница 4: ...into the BioSpot BT600 device It is prohibited to store or place any object on the platen of the BioSpot BT600 which is not in the purpose of use Keep your hands away from all moving elements during o...

Страница 5: ...a laboratory or industrial environment only the BioSpot BT600 is to be operated by skilled personnel who are familiar with the user s manual only The BioSpot BT600 workstation without installed compo...

Страница 6: ...e in outdoor environment Use the device indoors in clean and dry environment The installation base needs to offer a stationary even and robust surface It must resist the appearing acceleration forces...

Страница 7: ...igure 1 Photograph of the BioSpot BT600 workstation indicating the most important functional parts Dispenser connection Plug socket to connect a dispenser e g up to eight PipeJet dispensers Warning Hi...

Страница 8: ...mponents from the power supply except the internal PC and the power switch on the back side The emergency switch can be used to stop non intended movement of one of the axis or non intended dosage PLE...

Страница 9: ...re essential methods to improve or support the development of the desired dosing process or be applied to implement process control Possible accessories are listed and explained in the following TopVi...

Страница 10: ...control by correlation of generated signals Pressure controlled reservoir o guarantees constant pressure at the dosing nozzle o improves long term stability of the dosing process o enables big reservo...

Страница 11: ...socket 90 240V o Fuse 6A B C characteristic 3 Make sure the emergency switch is unlocked 4 Switch the backside power toggle switch to 1 on to start the BioSpot BT600 o the internal PC will start up au...

Страница 12: ...into the substrate holder Move reservoir holder down and fix it with the allen key Connect Tygon ND 100 65 to the intermediate reservoir using the black rubber adaptors and the Serto aspiration connec...

Страница 13: ...on of the device you can request for the original set of registry entry by BioFluidix 7 2 Interfaces Ethernet socket Typ RJ45 Transfer rate 10 100 1000Mbit s USB 2 0 USB connectors DVI connector Conne...

Страница 14: ...nsions w x l x h 550 mm x 636 6 mm x 539 2 mm Weight 32 kg depending on the individual configuration Environmental operation conditions Temperature 10 C 30 C Humidity 10 80 non condensing conditions 7...

Страница 15: ...ispensing the BioSpot BT600 workstation check for contaminations of hazardous or environmental and health threatened substances that may have been formed during use and remove them if required Dispose...

Страница 16: ...stions or comments please contact BioFluidix GmbH Engessertsra e 4a D 79108 Freiburg Germany phone 49 761 458938 0 fax 49 761 458938 99 e mail info biofluidix com Internet www biofluidix com Freiburg...

Отзывы: