GeneExplorer Thermal Cycler 

Instruction for Use

 

 

 

 

                    GE-96G, GE-384G, GE-4I, GE-48D, GE-48DS, GE-48DG, GE-96S 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

   

 

 

 

 

 

 

 

 

Note:

 

Please  understand  that  no  further  notice  will  be  given  for  any 

Hangzhou Bioer Technology Co., Ltd. 

 

Содержание GE-48DS

Страница 1: ...GeneExplorer Thermal Cycler Instruction for Use GE 96G GE 384G GE 4I GE 48D GE 48DS GE 48DG GE 96S Note Please understand that no further notice will be given for any Hangzhou Bioer Technology Co Ltd...

Страница 2: ...ghts reserved Without prior written approval from Hangzhou Bioer Technology Co Ltd any part of the document shall not be copied re produced or translated into other language File No BYQ56170000000ESM...

Страница 3: ...nsequences arising from user s failure for abiding by the following requirements Attention The instrument is Type I equipment in line with GB4793 1 standard with protection level of IP20 The instrumen...

Страница 4: ...urce such as close to water pool or water pipe and the place should be well ventilated and free from corrosive gas or high magnetic field interruption Don t place the instrument in moisty or dusty are...

Страница 5: ...has left the factory in a perfectly safe condition The instrument meets the requirements of the Regulation EU 2017 746 of the European Parliament and of the Council of 5 April 2017 on in vitro diagnos...

Страница 6: ...ring instrument operation or within a certain period after the program ends B Other symbols on the packaging Date of manufacture Indicates the date when the medical device was manufactured RoHS Restri...

Страница 7: ...cal devices Quality management systems Requirements for regulatory purposes If the instrument is installed in the required working environment there is no need to calibrate or correct the temperature...

Страница 8: ...f warranty Attention After opening the packaging box immediately check the objects contained therein based on the packing list In case of any object damage or missing please immediately contact the su...

Страница 9: ...AM 8 2 CHECK BEFORE TURNING ON THE INSTRUMENT 9 3 TURN ON THE INSTRUMENT 9 CHAPTER FOUR OPERATION INSTRUCTIONS 10 1 TURN ON THE INSTRUMENT 10 2 PCR PROGRAM EDITING 12 2 1 SETTING OF PROGRAM SECTION 13...

Страница 10: ...ler Instruction for Use Chapter One Introduction 2 FURTHER NOTICE WILL BE GIVEN FOR ANY MODIFICATION 25 APPENDIX 2 VALUE AND MEANING OF SET PARAMETERS 26 APPENDIX 3 PRODUCT EMC STATEMENT 28 APPENDIX 4...

Страница 11: ...cytology virology oncology genetic epidemiology forensic pathology and flora and fauna immunology it has accomplished outstanding achievements and become a milestone of contemporary molecular biology...

Страница 12: ...n continue to complete the program unfinished Hot lid can adjust height and pressure based on actual using condition to adapt to different reaction tubes and effectively prevent reagent evaporation an...

Страница 13: ...GeneExplorer Thermal Cycler Instruction for Use Chapter One Introduction 5 6 Service life of Thermal Cycler Service life of Thermal Cycler is 5 years from the date of instrument installation...

Страница 14: ...he above requirements Pay special attention to reliable grounding of the power line 2 Transport and storage condition Ambient temperature 20 C 55 C RH 80 3 Basic parameters The parameters of the modul...

Страница 15: ...imum number of editable sections is 10 maximum number of steps is 100 maximum number of loop is 99 available for secondary nesting gradient temperature range is 1 30 Ladder Temperature range 0 1 6 2 f...

Страница 16: ...miliar with the contents in the chapter before turning on the machine 1 Structural schematic diagram The structural parts of the instrument include heating module hot lid power switch module including...

Страница 17: ...nt Turn on the power switch switch to the position of and then the instrument will carry out self test which will take about 20s please wait patiently If the self test has discovered no problems the m...

Страница 18: ...upplier Attention If the sample volume of your experiment is less than the number of cone holes of the module please evenly distribute the test tubes to ensure stable press of the hot lid on the test...

Страница 19: ...Operation Instructions 11 Picture for GE 96G Picture for GE 48DS Picture of Model GE 384G GE 4I GE 96S would show the module type with 384G plate and 96V respectively Press the icon New Experiment to...

Страница 20: ...em Settings During operation instrument operation status is displayed on the main interface and click to enter the interface of operation monitoring When operation is stopped click to enter the interf...

Страница 21: ...and Add Click Insert a new section will be added before the section selected and lighting up Click Add a section will be added to the end of the program Click Delete to delete the section selected 2...

Страница 22: ...op On the interface of PCR program setting click Loop and it will prompt interface 1 1 select the starting and ending sections of the loop click Determine and it will prompt the setting box 1 2 for lo...

Страница 23: ...e increase or decrease for each loop changes the positive and negative of data Extended time time increase for each loop Speed ramp rate Gradient gradients temperature refers to the temperature differ...

Страница 24: ...le name to save the file Files saved can be found under Open Experiment 3 Run program Before running experiment set up hot lid temperature reaction volume temperature control mode and control mode on...

Страница 25: ...will display the temperature of each module 3 1 Temperature control mode Control Mode represents the default temperature control mode after the system is turn on and it includes two modes of Block and...

Страница 26: ...ach the set temperature more quickly If setting time equal to that of Block mode the actual time for reagent reaching set temperature is obviously extended see the figure below for the relation betwee...

Страница 27: ...rmal Cycler Operation Manual Chapter Four Operation Instructions 19 New Experiment Click to enter into user management Press Function to delete or rename user Select user to export to USB flash disk C...

Страница 28: ...structions 20 Press Function to delete or rename file Select file to export to USB flash disk 5 System settings Click System Settings on the main interface to enter the interface of setting which incl...

Страница 29: ...e access 3 Hot lid related 4 Time and date set instrument time and it will display on the lower right corner of the main interface 5 Instrument part use time 6 Dairy 7 About instruments 8 Tm calculati...

Страница 30: ...er Thermal Cycler Operation Manual Chapter Four Operation Instructions 22 6 2 Driver upgrade File Copy to USB flash disk GeneExplorer location Enter into System Settings about instruments select drive...

Страница 31: ...perature rise rate or imprecise temperature control Blocked ventilation holes Remove the blocking objects on ventilation holes Loose connecting line Open the shell and check the connecting line tighte...

Страница 32: ...low 0X Left sensor Right sensor Reserve d Reserve d Reserve d Reserve d Single module hot lid double modu le hot lid A Double modul e hot lid B Radiato r 0 normal 1 open circuit alarm 2 short circuit...

Страница 33: ...Gene Explorer Thermal Cycler Operation Manual Appendix 1 25 Appendix 1 Wiring Diagram the diagram is only for reference and please understand that no further notice will be given for any modification...

Страница 34: ...ime Time increase for each cycle m ss 0 00 9 59 6 Cycle1 Number of the cycle 1 5 Cycle1 00 From 00 to 00 00 Times of the cycle 1 99 From 00 Starting section of the cycle X any existing section number...

Страница 35: ...ut 3 after reaching set temperature reagent temperature can get close to the set temperature more quickly see Fig 2 13 Sample Volume Sample volume ul 10 200 14 Holid Hot lid off Off Hot lid on 30 110...

Страница 36: ...formance criterion Shell Electrostatic discharge ESD GB T 17626 2 Air discharge 2kV 4kV 8kV Contact discharge 2kV 4kV B Radiation magnetic field GB T 17626 3 3V m 80MHz 2 0GHz 80 AM A Rated PFMF a GB...

Страница 37: ...equipment with DC power input port using AC DC power adapter test should be conducted on AC input port of AC DC power adapter specified by the manufacturer If there is no specific requirement typical...

Страница 38: ...Model Specification Unit QTY Remark 1 Power line with shield magnetic ring clamp 250V10A String 1 Accessory 2 Fuse 250V12A 5 20 Piece 2 Accessory 3 WiFi Module Piece 1 Accessory 4 Ethernet Patch Cable...

Страница 39: ...g District 310053 Hangzhou PEOPLE S REPUBLIC OF CHINA Website www bioer com cn Tel 86 571 85800535 87774558 Fax 86 571 85800537 87774559 E mail instrument bioer com cn MedNet EC REP GmbH Borkstrasse 1...