To Access the Linearity Testing Menu:
1.
From the QA Dose Calibrator screen touch <Linearity Testing>. The Linearity Testing
screen is displayed.
Figure 6.43. The Linearity Testing Screen.
2.
Select a function from the menu by touching the desired test button.
Manual Decay Test
Figure 6.44. The Manual Decay Test Screen.
This test allows the user to record the linearity readings each time that <Record Reading> is
touched. The user can select the calculation method using a normalized reading or an
averaging method for the calculation. This choice is made as part of setting the test
parameters prior to starting the linearity test. The manual decay test can be paused to use
the dose calibrator for measuring functions and still return to the manual decay testing
operation.
To Select Setup Options:
1.
From the Manual Decay Test screen, touch <Set Test Parameters>. The Manual
Decay Test Parameters screen is displayed.