the message Test Saved at the bottom.
Figure 6.24. Geometry Test Screen with Test Saved Message
15.
Touch <Back> to return to the Dose Calibrator Quality Assurance menu.
16.
Touch <New Test> to begin a new Geometry Test.
Geometry Test History
Figure 6.25. The Geometry Test History Screen
The Geometry Test History screen displays the Geometry Tests performed on the dose
calibrator. The display shows the isotopes used for a Geometry Test. Any Geometry Test
outside the acceptable variance is shown in red.
•
Touch <Back> to return to the QA Dose Calibrator screen.
•
Touch <Print> to print the displayed Geometry Test History data.
To Delete Data:
1.
Use the <
▲
> and <
▼
> arrows to scroll to the test to delete.
2.
Touch <Delete>. The Delete Stored Test Data screen is displayed. Touch <Accept> to
delete the test and return to the Geometry Test History screen. Touch <Cancel> to return
to the Constancy Test History screen without deleting the test.