Figure 6.18. Select Isotope Screen
3.
Touch <Set Test Parameters>. The Set Test Parameters screen is now displayed.
Figure 6.19. The Geometry Set Test Parameters screen.
4.
Set the Container Description by touching the <Type>, <Size>, and <Material> icons.
<Type> adjusts syringe or vial, <Size> adjusts from 1 to 30 ml, <Material> is glass or
plastic.
5.
Touch to highlight volume 1 and use the <
▲
> and <
▼
> arrows to adjust the value.
Touch volume 2 and the system automatically increases by the appropriate volume.
Use the <
▲
> and <
▼
> arrows to adjust the volume. Touch volume 3 and the
system automatically increases by the appropriate volume. Use the <
▲
> and <
▼
>
arrows to adjust the value. Continue entering the volumes until the maximum
volume for which the system is to test is reached.
6.
Use the <
▲
> and <
▼
> arrows to set the normalization volume.