To Delete Data:
1.
Use the <
▲
> and <
▼
> arrows to scroll to the test to delete.
2.
Touch <Delete>. The Delete Stored Test Data screen is displayed. Touch <Accept>
to delete the test and return to the Constancy Test History screen. Touch <Cancel> to
return to the Constancy Test History screen without deleting the test.
Figure 6.11. Delete Stored Test Data Screen.
To View Details:
Figure 6.12. Constancy Test View Details Screen
The Constancy Test History screen displays the constancy tests performed on the dose
calibrator. The display shows all the isotopes for which constancy tests have been completed by
the date performed. Any constancy test outside the acceptable variance is shown in red. Daily
constancy activity results can be displayed by month in a plotted graph format.
•
Touch <Back> to return to the QA Dose Calibrator screen.