Figure 6.4.
Zero Background from Constancy Test Screen
4.
Place the selected sealed source into the dipper and into the detector. Allow the reading to
settle as illustrated in the current activity section of the screen.
5.
Touch <Record Reading> to enter the reading. The variance from the decayed
calculated standard is shown.
Figure 6.5.
Record Readings from Constancy Test Screen
6.
If <Back> is touched, a message that recorded data will be lost is displayed. Touch <Yes> to
return to the Daily Constancy Test screen without saving the data. Touch <Cancel> to return
to the Daily Constancy Test screen with the data recorded.
Figure 6.6.
Exiting Message Screen
7.
Touch <Save> to save the Daily Constancy Data. The Save Results screen is displayed.
8.
Touch <Staff Member>. The Staff Member List is displayed. Touch to highlight any