Figure 4.63. Label Printing Preferences Screen.
Touch <YES> to return to the Dose Calibration Configuration screen without updating
the Label Printing Preferences or <NO> to remain on the Label Printing Preferences screen.
Source Detection Threshold
Figure 4.64. Source Detection Threshold Screen
The Source Threshold value determines the system’s response time. Above the threshold,
response time is quick. Below the threshold, response time is significantly slower.
To Change The Source Threshold:
1.
From the Utilities menu, touch < Dose Calibrator Configuration >. The Dose Calibrator
Configuration screen is displayed.
2.
Touch <Source Detection Threshold>. The Source Detection Threshold screen is displayed
(see Figure 4.64).
3.
Touch the red threshold number of the isotope on the screen to display the following
screen: