4.
Touch <Calibration Activity>. The Calibration Activity screen is now displayed.
Figure 4.48. Calibration Activity Screen
Using the displayed keypad, enter the activity for the date of calibration for the source and
touch <Accept>. Once the calibration activity has been entered, the user can increase or
decrease the amount by touching the <
▲
> or <
▼
> keys.
5.
The Sealed Source screen is redisplayed containing the calibration activity. If necessary,
touch the appropriate radio button to toggle between µCi and mCi (MBq and GBq for
Becquerels.)
Figure 4.49. Sealed Source Screen with Sealed Source and Calibration Activity
Displayed
6.
Touch <Calibration Date>. The Calibration Date screen is displayed defaulted to the current
date.