Press <No> to keep the changes, followed by <Accept> on the Setup Button Isotope screen.
Press <Yes> to exit to the Isotope List screen without saving any changes.
NOTE:
Only new isotopes added to the system by the user can be removed. Factory isotopes
cannot be deleted.
Change Dial Values
Figure 4.37. The Change Dial Values Screen
This option is used to change the dial value of any isotope on the Isotope Lists. The entire list
can be scrolled through displaying the original dial value for all isotopes and the new dial value
for any that have been changed. New dial values are preceded by an * when viewed on the
Dose Calibrator Measurement screen after touching <Select Isotope>.
To Change An Isotope Dial Value:
1.
From the <Home> screen, select <Utilities>. The Utilities screen is now displayed.
2.
Select <Dose Calibrator Configuration>. The Dose Calibrator Configuration screen is now
displayed.
3.
Select <Isotope List>.
4.
Enter the passcode and touch <Accept>. The Isotope List screen is now displayed.
5.
Touch <Change Dial Values>.
6.
The Change Dial Values screen is displayed with the user-entered isotopes alphabetically
incorporated.