command with great care. If the display is in use performing a QA test, it can corrupt the
current running test.
Example:
ISO 43
: Sets the isotope Tc-99m for detector #2.
The following errors may be returned:
Error: Invalid Isotope Index #
: A number greater than the maximum isotope was specified.
Error: Measure Screen Not Active:
The Measure screen must be on view for this command to
work.
When the command is received, the newly selected isotope will be displayed and the activity
changed accordingly. However, if any screen other than the Measure screen is on view, the
isotope cannot be changed and the displayed screen is not affected.
Note:
Reference the Isotope Index # Table for programming.
MODEL
Displays the abbreviated model name as follows:
•
Atomlab 500 = 500
•
Atomlab 500Plus = 500W
•
Wipe Test Counter = WIPE.
PWD {1/0}
Sets the screen password override (1=Override, 0=Entry Required). If no parameter is provided,
the current state of the password override is displayed.
QADATA
Displays a summary of stored QA Data Records. For each QA data type, the allocated space and
used space is displayed (see table below):
QA Data
Space
Used
Reportable
Constancy
99999
12
12
Geometry
99999
1
1
Accuracy
99999
2
2
Decay Data
99999
1
1
Lineator
99999
1
1
Calicheck
99999
1
1
HV Cal
99999
1
1
Daily Cal
99999
5
5
Background
99999
2
2
Wipe Count
99999
1
1
Chi Square
99999
1
1
Wipe Const
99999
1
1
Содержание ATOMLAB 500
Страница 1: ...ATOMLAB 500 DOSE CALIBRATOR OPERATION AND SERVICE MANUAL 086 330 086 332 FN 17 355 9 17...
Страница 89: ...Figure 6 28 A Geometry Test Reports Screen...
Страница 96: ...Figure 6 39 Accuracy Test Reports Screen 2 Touch Summary History Figure 6 40 Accuracy Test Summary Report...
Страница 162: ...10 Parts and Assembly Instructions TBD...
Страница 211: ......
Страница 212: ......