Error: Measure Screen Not Active
– The Measure screen must be on view for this command to
work.
Error: No Available Chamber
– The dose calibrator detector cannot be found or is not available.
If the background command succeeds, it may be performed instantly taking effect on the very
next activity reading or it may take up to 30 readings. If the Zero Background does not instantly
occur, each reading from that detector will be preceded by:
~ Calculating background… and followed by CR/LF.
When this command is received, the Zero Background screen is displayed.
DCNT
Display the number of active, connected detectors. It should be 1 for an Atomlab 500 or Wipe
Test Counter, 2 for an Atomlab 500Plus, or 0 if no detectors were found to be connected.
DS {1/0}
Turns Data Streaming Mode ON=1, OFF=0. If no parameter is given, the current data streaming
mode is displayed. When ON, every time a reading is reported from a dose calibrator, the
isotope’s symbol, dial values, and activity reading is reported continuously until turned to OFF.
The report starts with a tilde (~) character and the fields are separated by a space. The activity
reading always has three decimal digits in the report followed by the units.
For example:
~Tc-99m 37.1 1404.767 µCi
.
DSTAT{#}
Display the detector status for all or a specific detector. If no parameter is given, it will default
to all detectors, while 1…n is the detector number.
The data returned for a dose calibrator is as follows and in this order: Detector ID, High
Voltage, Pressure, Temperature, Argon gas (grams).
The data returned for a Wipe Test detector is as follows and in this order: Detector ID, Current
High Voltage, Last Saved High Voltage, and Channel #Cs-137 peak was detected at.
If there are multiple detectors connected, an additional field is added: the detector number.
Example of the returned data for an Atomlab 500
Plus
:
2 347.1 252.6 24.9 7.127 1
14 1092.4 1092.4 52 2
DT
Displays the current Date/Time of the Atomlab software in the region format and time zone
configured, but always in a 24-hour format.
Example in US format:
9/11/2017 12:53:00
ECHO {1/0}
Display or turn the local character echo ON=1 or OFF=0. The local character echo is on by
default. If data streaming is used, this should be turned OFF to avoid mangling the streaming
reports.
Содержание ATOMLAB 500
Страница 1: ...ATOMLAB 500 DOSE CALIBRATOR OPERATION AND SERVICE MANUAL 086 330 086 332 FN 17 355 9 17...
Страница 89: ...Figure 6 28 A Geometry Test Reports Screen...
Страница 96: ...Figure 6 39 Accuracy Test Reports Screen 2 Touch Summary History Figure 6 40 Accuracy Test Summary Report...
Страница 162: ...10 Parts and Assembly Instructions TBD...
Страница 211: ......
Страница 212: ......