Figure 6.75. Calicheck Initial Test Screen – Test Complete.
7.
Touch <Save>.
8.
If the appropriate staff member is not displayed, touch <Staff>. Touch any line on
the display to highlight the staff member. Touch <Accept> to return to the Save Results
screen.
9.
Touch <Accept> to select the staff member, save the data, and return to the Calicheck
Initial Test screen.
10.
Touch <Save Data> to save the data.
11.
Touch <New Test> to perform the test again.
12.
Touch <Measure>to advance to the Dose Calibrator Measure screen.
13.
Touch <Back> to return to the Lineator Testing screen.
Calicheck Test
Figure 6.76.
The Calicheck Test Screen.
The Calicheck Test is performed with this dose calibrator after the initial calibration factors
for the Calicheck function have been taken and saved. The current Calicheck calibration
test is a calculation of the Calicheck calibration factors with the current activity readings.
The Calicheck Linearity test should be performed at installation and at least quarterly