Figure 6.71. Lineator Test after Saving the Data.
11.
Touch <Measure>to advance to the Dose Calibrator Measure screen.
12.
Touch <Back> to return to the Lineator Testing screen.
NOTE:
The Lineator Test must be performed in less than six minutes or the test must be
redone because of decay. See the Lineator instructions.
Calicheck Initial Test
(Also Known As Calicheck Calibration Factor)
Figure 6.72. The Calicheck Initial Test Screen.
The first time the Calicheck is used with Atomlab 500, calibrator factors for the Calicheck
function must be set and the Atomlab 500 must be calibrated for the specific Calicheck
package using the initial procedure described below. The Atomlab 500 will store the initial
factors and print them in the factor column of the results table.
The initial Calicheck Test needs to be performed only once unless the Calicheck is
replaced or damaged, in which case the initial factors will need to be redone.