Figure 6.69. Lineator Test – Record Reading Touched.
5.
Touch <Record Reading> to record the reading for Tube 1 & 2.
6.
Continue taking readings by touching <Record Reading> after each new tube or set of
tubes is placed in the detector.
7.
Touch <End Test> to end the test at any time.
Figure 6.70. Lineator Test – End Test Touched.
8.
Touch <Save > to save the test just performed.
9.
If the appropriate staff member is not displayed, touch <Staff>. Touch any line on the
display to highlight the staff member. If a staff member has previously been
selected, the system will default to that person. Touch <Accept> to select the staff
member, save the data, and return to the Lineator Test screen.
10.
Touch <New Test> to perform the test again.