5.
Touch <Record Reading> to record the reading for Tube 1 & 2.
6.
Continue taking readings by touching <Record Reading> after each new tube or set of
tubes is placed in the detector.
7.
Touch <End Test> to end the test at any time.
Figure 6.66. Lineator Initial Test – End Test Touched
8.
Touch <Save> to save the test just performed. If the appropriate staff member
is not displayed, touch <Staff Member>. Touch any line on the display to
highlight the staff member. If a staff member has previously been selected, the
system will default to that person. Touch <Accept> to save the data in memory
and return to the Lineator Initial Test data screen.
9.
Touch <Measure> to advance to the Measure screen.
10.
Touch <New Test> to run another Lineator Test.
11.
Touch <Back> to return to the Lineator Testing screen.
Lineator Test
Figure 6.67. The Lineator Test Screen.
The Lineator Test is performed with this dose calibrator after the initial factors for the
Lineator function have been taken and saved. The Lineator Test is a comparison of the
initial factors with the current factors.