Figure 6.62. The Save Automated Decay Test Screen.
14.
The user can highlight to change what will be used for the normalized reading by touching
to highlight a different reading. Touch <Accept> to accept the reading and return to the
Automated Decay Test screen. The screen displays a “Test Complete” message. If <Back> is
touched, a prompt is displayed noting that recorded data will be lost. Touch <Yes> to
discard the data. Touch <No> to return to the test ended screen.
15.
Touch <Save > to save the data. If the appropriate staff member is not displayed,
touch <Staff>. Touch any line on the display to highlight the staff member. Touch
<Accept> to return to the Save Results screen.
16.
Touch <Accept> to select the staff member, save the data, and return to the Semi-
Automated Decay Test screen.
17.
Touch <New Test> to perform a new Semi-Automated Decay Test.
18.
Touch <Measure>to advance to the Dose Calibrator Measure screen.
19.
Touch <Back> to return to the Semi-Automated Decay Test screen.
20.
Touch <Back> a second time to return to the Linearity Testing screen.
Linearity Automated Decay Test
WARNING:
There are several important items to know when using the Decay
Tests.
1.
There are a maximum of 60 readings for the Decay Tests. The test will stop after the
60th reading, regardless of the activity reading. It will also stop when the reading drops
below 10 µCi.
WARNING:
There are several important items to know when using the
Automated Decay Test.
2.
The Automated Linearity Test will end automatically when the activity reading falls
below 10 µCi.
3.
If a normalized reading that occurs after the 10 µCi reading is selected, the system will
automatically select the middle reading as the normalized value.
4.
The Automated Decay Test displays the time to complete the test in the upper left of the
screen. This calculation is based on the half-life of the isotope being used to perform the
test and the activity reading from the first measurement.
Содержание ATOMLAB 500
Страница 1: ...ATOMLAB 500 DOSE CALIBRATOR OPERATION AND SERVICE MANUAL 086 330 086 332 FN 17 355 9 17...
Страница 89: ...Figure 6 28 A Geometry Test Reports Screen...
Страница 96: ...Figure 6 39 Accuracy Test Reports Screen 2 Touch Summary History Figure 6 40 Accuracy Test Summary Report...
Страница 162: ...10 Parts and Assembly Instructions TBD...
Страница 211: ......
Страница 212: ......