Figure 6.60. Automated Decay Test Screen displaying Readings.
12.
Touch <Measure> to advance to the Measure screen and use the system as a standard
dose calibrator. Touch <Return to Decay Test> on the Measure screen to return to the
automated decay test. Touch <Resume Test> to resume the Automated Decay Test.
NOTE:
Remember to place the source back into the detector before selecting <Resume
Test>.
Figure 6.61. The Measuring screen during <Pause Test>
.
NOTE:
If the time for the reading has passed, the system will take a reading and record the
time once the test is resumed. It does not skip a reading that was not taken at the correct
time.
13.
Touch <End Test> once enough decay points have been recorded to end the test. The
system automatically takes a final reading and ends the test when the activity is below 10
µCi. Once the test has ended, the system allows the user to <Select Normalized Reading>
and change the reading selected as the normalized value. The Count time measured closest
to the normalized reading time will be highlighted to be used as the normalized reading.