Installation Instructions 

Tools Required  

1.

 

Allen wrench set, 1/8" 5/16" 

2.

 

Socket set with extension, 1/2", 9/16", 7/1", 3/4", 5/8" 

3.

 

Open end or closed end wrench, 1/2", 9/16", 7/1", 3/4", 5/8" 

4.

 

Flat head and Phillips head screwdriver 

5.

 

Tin snips 

6.

 

Razor blade 

Installation Procedure  

1.

 

Uncrate contents from the boxes. 

2.

 

Inspect for any signs of damage. 

3.

 

Place the chair T-base (larger T-base) in the area of choice. Be aware that the chair 
must recline and that, for the Pro model, the chair is 220VAC actuated. 

4.

 

Place the dynamometer T-base (smaller T-base) approximately 6" in front of the 
chair T-base (do not align). 

5.

 

With a 1/8" Allen wrench, remove the screws from the chair T-base cover plate. 

6.

 

Remove the smaller bolts (1/2" socket). 

7.

 

Remove the larger bolts (3/4" socket). 

8.

 

Slide the dynamometer T-base into place and align the holes. 

9.

 

Install the T-Base bolts that go in toward the floor with a 1/2" socket. 

NOTE:

 It may be necessary to keep the bolts loose until the bolts going into the side of the 

dynamometer T-Base are aligned. Make certain to go back and tighten the bolts after the 
front T-Base bolts are installed. 

10.

 

Tighten the bolts going into the side of the T-base with the 3/4" socket. 

11.

 

Remove the chair from the pallet with a 5/16" Allen wrench. 

12.

 

Place the chair on top of the chair T-base. Align the holes and use a 5/16" Allen 
wrench to tighten. Be careful not to slide the chair around too much as this may 
cause the paint to scratch. 

13.

 

Remove the dynamometer and gimbal from the pallet with a 5/16" Allen wrench. 

14.

 

Place the dynamometer and gimbal on the T-base and tighten with a 5/16" Allen 
wrench. Ensure that the power cords are facing forward (away from the chair). 

15.

 

Unpack the computer, monitor, printer, keyboard, mouse, mouse pad, and hand-
held Comfort Stop. 

16.

 

Plug the pause switch into the connector located on the CDS midway up on the 
right side. 

17.

 

Connect the Comfort Stop to the dynamometer. 

18.

 

Install the monitor (refer to the Monitor Install Procedure.) The power and signal 
cable are already run and connect to the monitor. 

19.

 

Install the computer, located on the bottom shelf in the front of the CDS. Remove 
the side covers (lift up and pull out and down) and slide the back of the computer 
in from the left side. The cables are already run and ready to plug into the 
computer as labeled. 

20.

 

Place the printer on the middle shelf. The cables are already run and located on the 
right side of the shelf. 

21.

 

Plug in the power cord and connect the printer cable into the back of the printer. 

22.

 

Install the Motor Power cord and the Dyna Data cord into their respective, labeled, 
connectors at the rear of the CDS cart. 

23.

 

Install cabling at the back of the T-Base (refer to AC Power Installation procedure.) 

Содержание 840-000,

Страница 1: ...BIODEX MULTI JOINT SYSTEM INSTALLATION INSTRUCTIONS 840 000 840 000 10 840 000 30 850 000 850 000 10 850 000 30 FN 08 051 CLR Rev A 9 19...

Страница 2: ...request or directly from the Biodex website http www biodex com If the desired information is not found please feel free to contact a local distributor or Biodex directly at supportservices biodex com...

Страница 3: ...mbols 4 1 Installation Instructions 6 Tools Required 6 Installation Procedure 6 Monitor Install Procedure 8 Cabling for CDS Power Receptacle 8 Seat Lift Power Connection 10 AC Input Receptacle Power C...

Страница 4: ...4 Definition of Symbols The following symbols and their associated definitions are used and implied throughout this manual...

Страница 5: ...ns System Power Requirements Provide 220 VAC 60 Hz 20 amp dedicated electrical service with dedicated ground Locate within ten feet of ear of the T base Receptacle will be a single Hubbell IG 5461 rat...

Страница 6: ...he chair on top of the chair T base Align the holes and use a 5 16 Allen wrench to tighten Be careful not to slide the chair around too much as this may cause the paint to scratch 13 Remove the dynamo...

Страница 7: ...7 Figure 1 1 Multi Joint System Cart and Monitor from the Rear...

Страница 8: ...cedure Figure 1 2 Rear View of the CDS Cart Cabling for CDS Power Receptacle Figure 1 3 Power Receptacles on the Back of Seat CDS Cart Power Receptacle Seat Lift Power Receptacle 195 to 245 Volt AC In...

Страница 9: ...e Screws from the CDS Cart Power Receptacle Figure 1 5 Loosen the Clamp on the Cable Figure 1 6 Cable Positioned at CDS Cart Power Receptacle 1 Strain Relief Slot Facing Up Cable Clamp Strain Relief S...

Страница 10: ...Cart Power Receptacle Seat Lift Power Connection Figure 1 8 Loosen the Cable Clamp Screw and Replace the Clamp on the Cable Figure 1 9 Seat Lift Receptacle 1 Seat Lift Receptacle 1 2 3 Strain Relief...

Страница 11: ...t Lift Power Cord AC Input Receptacle Power Connection Figure 1 11 Loosen the Clamp on the Cable Figure 1 12 Loosen the Cable Clamp Screw and Remove the Strain Relief Cable Clamp Screw Seat Lift Power...

Страница 12: ...12 Figure 1 13 Removing the Screws from the AC Input Receptacle Figure 1 14 Installing the AC Input Cable 3 3 1 2 Cable Clamp Screw...

Страница 13: ...STATE CONTACT AT SITE PHONE FAX Biodex Representative Installation Start Time Serial Number Model Number Software Version Firmware Version Biodex Contacts 1 800 224 6339 Software Support ext 2120 Tech...

Страница 14: ...pauses both on the dynamometer and hand held and rotate buttons on the dynamometer Comments 6 In the reactive eccentric mode check speeds and torque limits Comments 7 In the isokinetic mode check spee...

Страница 15: ...nctioning properly The Biodex representative has instructed the facility on startup hardware and software The facility is capable of proper operation of the Biodex System BIODEX REPRESENTATIVE DATE Si...

Страница 16: ...16...

Отзывы: