3. Click on the <Preferences> button and then enter a total urine volume for the patient that you are
testing. Once you have entered the total volume, click <OK>. You can now proceed with counting.
NOTE: You can change the sample volume, correction factor and count time for this study in the parameters
window.
NOTE: If you use a different standard volume or concentration, you must change the correction factor.
4. Click on the background button and the system will proceed to count lab background. When it is
done counting, the number of counts and the counts per minute will both be displayed.
5. Remove the background vial and insert the Co-57 standard into the well.
6. Click on the highlighted <Co-57> standard button. The system will now count the Co-57 standard.
When counting is finished, the total counts and the counts per minute will be displayed. The
<Sample> button should now be displayed.
7. Remove the Co-57 standard from the well and place the urine sample into the well.
8. Click on <Sample> to begin counting the urine sample. When counting is finished, the total counts
and counts per minute will be displayed.
9. When the system has completed the count, a window will be displayed asking to verify that the test
is complete. To save the results, click <Yes>. If you do not wish to Save, click <No> and you can
then proceed to recount any one of the three items counted.
NOTE: If an error was made in the total urine volume, do not click <Save>. You can go back and click on
<Preferences> to correct the error as long as you have not saved.
10. Once you have finished with any recounts for this patient, select <Save> to display the "Verify Test
Complete" screen. If the numbers are satisfactory, click <Yes> to save the results.
11. To print results at any time, click on the <Print> icon. This will open the report generation screen
from which you can choose a technologist to add to your study.
12. To enter and store the dose activity associated with this patient, press <Preview> for a preview of the
report. Press <Print> to print the report.
NOTE: Make sure to save before leaving the patient and going on to do other tests or studies with your
spectrometer.
NOTE: The Bracco and Mallinckrodt kits use the same procedure. You can modify the sample volumes, but
then you must correct and enter the appropriate sample volume in milliliters in the Preferences window.
Schilling Test for the Simplified Mallinckrodt Procedure
(See Figure 8-3.)
This is the same as the Mallinckrodt and Bracco procedures except that you cannot enter a sample
volume in the Preferences window. For this test, you must use a 4ml volume in order for the system to
calculate correctly.
Schilling Dicopac
(See Figures 8-4 and 8-5.)
The Dicopac Schillings procedure is programmed to use the Mediphysics/Amersham Dicopac Schilling
Test Kit. The counting procedures and formulas for calculations are all from the manufacturer's Dicopac
Kit instructions.
CONTENTS
8-5
SCHILLING TESTS
Содержание 187-130
Страница 8: ...1 INTRODUCTION SETUP INSTRUCTIONS 2 2 Figure 2 1 The MCA Uptake Stand ...
Страница 10: ...1 INTRODUCTION SETUP INSTRUCTIONS 2 4 Figure 2 2 The MCA Uptake Stand side view ...
Страница 20: ...1 INTRODUCTION SETUP INSTRUCTIONS 2 14 Figure 2 6 The System Setup Screen ...
Страница 22: ...1 INTRODUCTION SETUP INSTRUCTIONS 2 16 Figure 2 7 The Edit Physicians and Technologists Information Screen ...
Страница 25: ...1 INTRODUCTION 2 19 SETUP INSTRUCTIONS Figure 2 8 The Database Manager Screen ...
Страница 28: ...CONTENTS 3 3 ADMINISTRATION Figure 3 1 The Calibration Screen ...
Страница 29: ...1 INTRODUCTION Figure 3 2 A Calibration Report ADMINISTRATION 3 4 ...
Страница 31: ...1 INTRODUCTION Figure 3 3 The High Voltage Adjustment Screen ADMINISTRATION 3 6 ...
Страница 32: ...1 INTRODUCTION Figure 3 4 The High Voltage Report 3 7 ADMINISTRATION ...
Страница 34: ...1 INTRODUCTION Figure 3 5 The Chi Square Screen 3 9 ADMINISTRATION ...
Страница 35: ...1 INTRODUCTION Figure 3 6 The Chi Square Report ADMINISTRATION 3 10 ...
Страница 37: ...1 INTRODUCTION Figure 3 7 An Administration Report ADMINISTRATION 3 12 ...
Страница 39: ...ADMINISTRATION 3 14 1 INTRODUCTION Figure 3 8 The Isotope Editing Screen ...
Страница 40: ...1 INTRODUCTION 3 15 ADMINISTRATION Figure 3 9 The Select Isotope Window ...
Страница 42: ...1 INTRODUCTION 3 17 ADMINISTRATION Figure 3 10 The Detector Efficiency Screen ...
Страница 44: ...1 INTRODUCTION Figure 3 11 The Geometric Efficiency Window 3 19 ADMINISTRATION ...
Страница 48: ...1 INTRODUCTION 3 23 ADMINISTRATION Figure 3 13 A Spectrum Analysis Report ...
Страница 50: ...1 INTRODUCTION Figure 4 1 The Procedure Definition Screen THYROID UPTAKE 4 2 ...
Страница 51: ...1 INTRODUCTION 4 3 THYROID UPTAKE Figure 4 2 The Patient Definition Screen ...
Страница 56: ...THYROID UPTAKE 4 8 1 INTRODUCTION Figure 4 3 The Thyroid Uptake Screen ...
Страница 58: ...THYROID UPTAKE 4 10 1 INTRODUCTION Figure 4 4 A Thyroid Uptake Report ...
Страница 59: ...1 INTRODUCTION Figure 4 5 A Time Activity Report 4 11 THYROID UPTAKE ...
Страница 63: ...MANUAL MODE 5 2 1 INTRODUCTION Figure 5 1 The Manual Mode Screen ...
Страница 64: ...1 INTRODUCTION 5 3 MANUAL MODE Figure 5 2 A Manual Mode Report ...
Страница 66: ...1 INTRODUCTION WIPE TEST 6 2 Figure 6 1 The Wipe Counting Screen ...
Страница 67: ...6 3 WIPE TEST CONTENTS Figure 6 2 The Wipe Area Setup Screen ...
Страница 68: ...CONTENTS WIPE TEST 6 4 Figure 6 3 The Wipe Preferences Window ...
Страница 70: ...1 INTRODUCTION Figure 6 4 The Background LLD Screen WIPE TEST 6 6 ...
Страница 72: ...1 INTRODUCTION WIPE TEST 6 8 Figure 6 5 The Wipe Results Report ...
Страница 76: ...1 INTRODUCTION 7 3 BIOASSAY Figure 7 2 The Bioassay Preferences Screen ...
Страница 78: ...7 5 BIOASSAY CONTENTS Figure 7 3 An Individual Bioassay Report ...
Страница 79: ...CONTENTS Figure 7 4 An Employee Bioassay Summary BIOASSAY 7 6 ...
Страница 80: ...1 INTRODUCTION 7 7 BIOASSAY Figure 7 5 An Individual Bioassay History Summary ...
Страница 82: ...1 INTRODUCTION SCHILLING TESTS 8 2 Figure 8 1 Defining a Facility Standard Schilling Test ...
Страница 83: ...CONTENTS 8 3 SCHILLING TESTS Figure 8 2 The Schillings Counting Page ...
Страница 84: ...CONTENTS SCHILLING TESTS 8 4 Figure 8 3 A sample Schillings Mallinckrodt Report ...
Страница 87: ...CONTENTS 8 7 SCHILLING TESTS Figure 8 4 The Diopac Schillings Counting Page ...
Страница 88: ...SCHILLING TESTS 8 8 CONTENTS Figure 8 5 A Sample Diopac Schillings Report ...
Страница 90: ...CONTENTS HEMATOLOGY 9 2 9 1 The IHSA Counting Page ...
Страница 92: ...1 INTRODUCTION HEMATOLOGY 9 4 Figure 9 2 A Sample IHSA I 125 Test Report ...
Страница 93: ...CONTENTS 9 5 HEMATOLOGY Figure 9 3 The CR 51 Counting Page ...
Страница 95: ...CONTENTS 9 7 HEMATOLOGY Figure 9 4 A Sample Cr 51 Volume Test Report ...
Страница 96: ...CONTENTS Figure 9 5 The Cr 51 Red Blood Cell Survival Counting Page HEMATOLOGY 9 8 ...
Страница 98: ...CONTENTS HEMATOLOGY 9 10 Figure 9 6 A sample Cr 51 Survival Test Report ...
Страница 100: ...CONTENTS Figure 9 7 The GFR Counting Page HEMATOLOGY 9 12 ...
Страница 103: ...1 INTRODUCTION 9 15 HEMATOLOGY Figure 9 8 A One Sample Glomerular Filtration Rate Report ...
Страница 104: ...1 INTRODUCTION HEMATOLOGY 9 16 Figure 9 9 A Two Sample Glomerular Filtration Rate Report ...
Страница 106: ...1 INTRODUCTION HEMATOLOGY 9 18 Figure 9 10 The ERPF Counting Page ...
Страница 108: ...1 INTRODUCTION HEMATOLOGY 9 20 Figure 9 10 The ERPF Counting Page ...
Страница 109: ...1 INTRODUCTION 9 21 HEMATOLOGY Figure 9 11 An Effective Renal Plasma Flow Report ...