5.
To find the peak in the ROI, click on <Find Peak>. Click on <Save> to save this second region of
interest as ROI #2. To create additional regions of interest, click on <Show Cursors> and repeat steps
4 and 5.
6.
To print a Spectrum Analysis Report, click on <Print Preview> at the bottom right of the Cursor
Window for a preview. To print the report, click on <Print>.
7.
To exit Spectrum Analysis, click on <Close> at the bottom center of the Spectrum Analysis screen.
How is ROI Determined?
The preset ROI energy window values do not change as the user switches between well or probe,
regardless of the mode selected (i.e., Uptake, Schilling, MCA, etc.). The ROI is defined in units of energy
(keV) and then converted to channel numbers with the keV/channel slope value that is measured during
calibration. The channel number for the ROI values can and will change from probe to well if the
calibration keV/channel is different between probe and well.
Example
Tc-99m has a peak energy of 140 keV according to the radionuclide decay scheme. The Atomlab 950 uses
a default of +/- 15% energy window around the peak energy. These values calculate to be 119 keV to 162
keV and are displayed on the Isotope Editing page. When either the probe or well is selected, the ROI
channels are calculated based upon the Cs-137 calibration results for that particular detector. If the
calculation of the probe results in a .3775 keV/channel on gain 8, the TC-99m default gain, the channel
ROI will be 119 keV/.3775 keV channel = channel 315 and 162/.3775 = channel 429. For the well detector,
we might have a calibration of .3325 keV/channel. In this case, the ROI channels will calculate channel
358 to 487. Even though the ROI channels are different, the energy windows are the same. This will
become apparent when the spectrum for each detector is printed.
NOTE: It is not necessary to fix a symmetrical percent around the peak. In fact, some isotopes are multi-peaked,
which can cause an asymmetric spectrum. Keep in mind that if a calibration is performed you must save the result
in order to have it used in ROI calculations.
If you select an isotope from the Isotope Editing page, the screen will display the isotope name, LLD,
ULD, and gain setting as demonstrated in the following chart.
Gain
Max keV
Half Scale
1
3000
1500
2
1500
750
4
750
375
8
375
187
12
250
125
48
64
32
A.
Lower Level Discriminator (LLD) = 0.85 times the Min keV of interest.
B.
Upper Level Discriminator (ULD) = 1.15 times the Max keV of interest.
1. INTRODUCTION
ADMINISTRATION
3-22
Содержание 187-130
Страница 8: ...1 INTRODUCTION SETUP INSTRUCTIONS 2 2 Figure 2 1 The MCA Uptake Stand ...
Страница 10: ...1 INTRODUCTION SETUP INSTRUCTIONS 2 4 Figure 2 2 The MCA Uptake Stand side view ...
Страница 20: ...1 INTRODUCTION SETUP INSTRUCTIONS 2 14 Figure 2 6 The System Setup Screen ...
Страница 22: ...1 INTRODUCTION SETUP INSTRUCTIONS 2 16 Figure 2 7 The Edit Physicians and Technologists Information Screen ...
Страница 25: ...1 INTRODUCTION 2 19 SETUP INSTRUCTIONS Figure 2 8 The Database Manager Screen ...
Страница 28: ...CONTENTS 3 3 ADMINISTRATION Figure 3 1 The Calibration Screen ...
Страница 29: ...1 INTRODUCTION Figure 3 2 A Calibration Report ADMINISTRATION 3 4 ...
Страница 31: ...1 INTRODUCTION Figure 3 3 The High Voltage Adjustment Screen ADMINISTRATION 3 6 ...
Страница 32: ...1 INTRODUCTION Figure 3 4 The High Voltage Report 3 7 ADMINISTRATION ...
Страница 34: ...1 INTRODUCTION Figure 3 5 The Chi Square Screen 3 9 ADMINISTRATION ...
Страница 35: ...1 INTRODUCTION Figure 3 6 The Chi Square Report ADMINISTRATION 3 10 ...
Страница 37: ...1 INTRODUCTION Figure 3 7 An Administration Report ADMINISTRATION 3 12 ...
Страница 39: ...ADMINISTRATION 3 14 1 INTRODUCTION Figure 3 8 The Isotope Editing Screen ...
Страница 40: ...1 INTRODUCTION 3 15 ADMINISTRATION Figure 3 9 The Select Isotope Window ...
Страница 42: ...1 INTRODUCTION 3 17 ADMINISTRATION Figure 3 10 The Detector Efficiency Screen ...
Страница 44: ...1 INTRODUCTION Figure 3 11 The Geometric Efficiency Window 3 19 ADMINISTRATION ...
Страница 48: ...1 INTRODUCTION 3 23 ADMINISTRATION Figure 3 13 A Spectrum Analysis Report ...
Страница 50: ...1 INTRODUCTION Figure 4 1 The Procedure Definition Screen THYROID UPTAKE 4 2 ...
Страница 51: ...1 INTRODUCTION 4 3 THYROID UPTAKE Figure 4 2 The Patient Definition Screen ...
Страница 56: ...THYROID UPTAKE 4 8 1 INTRODUCTION Figure 4 3 The Thyroid Uptake Screen ...
Страница 58: ...THYROID UPTAKE 4 10 1 INTRODUCTION Figure 4 4 A Thyroid Uptake Report ...
Страница 59: ...1 INTRODUCTION Figure 4 5 A Time Activity Report 4 11 THYROID UPTAKE ...
Страница 63: ...MANUAL MODE 5 2 1 INTRODUCTION Figure 5 1 The Manual Mode Screen ...
Страница 64: ...1 INTRODUCTION 5 3 MANUAL MODE Figure 5 2 A Manual Mode Report ...
Страница 66: ...1 INTRODUCTION WIPE TEST 6 2 Figure 6 1 The Wipe Counting Screen ...
Страница 67: ...6 3 WIPE TEST CONTENTS Figure 6 2 The Wipe Area Setup Screen ...
Страница 68: ...CONTENTS WIPE TEST 6 4 Figure 6 3 The Wipe Preferences Window ...
Страница 70: ...1 INTRODUCTION Figure 6 4 The Background LLD Screen WIPE TEST 6 6 ...
Страница 72: ...1 INTRODUCTION WIPE TEST 6 8 Figure 6 5 The Wipe Results Report ...
Страница 76: ...1 INTRODUCTION 7 3 BIOASSAY Figure 7 2 The Bioassay Preferences Screen ...
Страница 78: ...7 5 BIOASSAY CONTENTS Figure 7 3 An Individual Bioassay Report ...
Страница 79: ...CONTENTS Figure 7 4 An Employee Bioassay Summary BIOASSAY 7 6 ...
Страница 80: ...1 INTRODUCTION 7 7 BIOASSAY Figure 7 5 An Individual Bioassay History Summary ...
Страница 82: ...1 INTRODUCTION SCHILLING TESTS 8 2 Figure 8 1 Defining a Facility Standard Schilling Test ...
Страница 83: ...CONTENTS 8 3 SCHILLING TESTS Figure 8 2 The Schillings Counting Page ...
Страница 84: ...CONTENTS SCHILLING TESTS 8 4 Figure 8 3 A sample Schillings Mallinckrodt Report ...
Страница 87: ...CONTENTS 8 7 SCHILLING TESTS Figure 8 4 The Diopac Schillings Counting Page ...
Страница 88: ...SCHILLING TESTS 8 8 CONTENTS Figure 8 5 A Sample Diopac Schillings Report ...
Страница 90: ...CONTENTS HEMATOLOGY 9 2 9 1 The IHSA Counting Page ...
Страница 92: ...1 INTRODUCTION HEMATOLOGY 9 4 Figure 9 2 A Sample IHSA I 125 Test Report ...
Страница 93: ...CONTENTS 9 5 HEMATOLOGY Figure 9 3 The CR 51 Counting Page ...
Страница 95: ...CONTENTS 9 7 HEMATOLOGY Figure 9 4 A Sample Cr 51 Volume Test Report ...
Страница 96: ...CONTENTS Figure 9 5 The Cr 51 Red Blood Cell Survival Counting Page HEMATOLOGY 9 8 ...
Страница 98: ...CONTENTS HEMATOLOGY 9 10 Figure 9 6 A sample Cr 51 Survival Test Report ...
Страница 100: ...CONTENTS Figure 9 7 The GFR Counting Page HEMATOLOGY 9 12 ...
Страница 103: ...1 INTRODUCTION 9 15 HEMATOLOGY Figure 9 8 A One Sample Glomerular Filtration Rate Report ...
Страница 104: ...1 INTRODUCTION HEMATOLOGY 9 16 Figure 9 9 A Two Sample Glomerular Filtration Rate Report ...
Страница 106: ...1 INTRODUCTION HEMATOLOGY 9 18 Figure 9 10 The ERPF Counting Page ...
Страница 108: ...1 INTRODUCTION HEMATOLOGY 9 20 Figure 9 10 The ERPF Counting Page ...
Страница 109: ...1 INTRODUCTION 9 21 HEMATOLOGY Figure 9 11 An Effective Renal Plasma Flow Report ...