
Before using the Atomlab 950 PC for counting purposes, it is important that you take a few minutes and
prepare the system for operation. The following functions are set from the Operation screen and control
the system throughout the different program modes.
NOTE: Before you begin with this chapter, be sure to enter your facility name and address in the Site Information
area, as explained in the section on System Setup. This will allow the system to print the facility name on all
printouts
•
Daily Calibration
•
Chi-Square Test
•
High Voltage Adjustment
•
Isotope Editing
•
Isotope Efficiencies
•
Detector Efficiency
•
Analysis (also available for each operation)
•
MCA Operation
Daily Calibration
(See Figure 3-1and 3-2.)
NOTE: The system must be ON for at least one hour before performing a calibration. This will allow time for the
unit to warm-up and stabilize.
Calibration Information
A daily system calibration must be performed each day the Atomlab 950 is to be used. Separate
calibrations are performed for the probe and well. The system keeps an internal listing of its last
calibration date and displays a message that calibration is needed. The user must then either perform a
calibration or click on the prompt that a calibration will not be performed. The system will not allow
measurement procedures to commence until the calibration prompt has been answered. Standard
calibration may be performed from any operating mode.
Calibration is performed automatically from the MCA Calibration screen. This screen advises when the
last high voltage and calibration were performed for your detectors. It also checks to ensure that the MCA
box and CPU are connected and communicating with each other.
Additionally, the calibration calculates full width half max (FWHM). FWHM is determined by
interpolating between two immediate points above and below the 50% value of the spectrum peak on
both sides of the peak. The difference between the two interpolated channel numbers are then divided by
the channel of the peak itself and multiplied by 100% to arrive at FWHM. The peak is simply the
maximum value in the spectrum.
It is not required to calibrate both detectors each day unless you are going to use them. If you do not
calibrate the detector when you go to use that detector, you will get a message advising you to calibrate.
It is recommended to do your calibrations in the mornings daily so that if you need to use the system, it is
ready to go.
If desired, you may perform additional calibrations on the same day whenever you desire by clicking on
the <Calibration> mode tab.
While the daily calibration is counting, in the upper right you see the ROI counts incrementing and the
CPM incrementing as well. The elapsed time is shown next to the time button. Below the start button is a
box listing estimated time left. This gives the remaining time to complete the calibration. The spectrum
increments as you count as the packets with the data are downloaded to the computer from the MCA.
The <Abort> button is highlighted and is the default button used in this mode at this point. You will
3. ADMINISTRATION
3-1
ADMINISTRATION
Содержание 187-130
Страница 8: ...1 INTRODUCTION SETUP INSTRUCTIONS 2 2 Figure 2 1 The MCA Uptake Stand ...
Страница 10: ...1 INTRODUCTION SETUP INSTRUCTIONS 2 4 Figure 2 2 The MCA Uptake Stand side view ...
Страница 20: ...1 INTRODUCTION SETUP INSTRUCTIONS 2 14 Figure 2 6 The System Setup Screen ...
Страница 22: ...1 INTRODUCTION SETUP INSTRUCTIONS 2 16 Figure 2 7 The Edit Physicians and Technologists Information Screen ...
Страница 25: ...1 INTRODUCTION 2 19 SETUP INSTRUCTIONS Figure 2 8 The Database Manager Screen ...
Страница 28: ...CONTENTS 3 3 ADMINISTRATION Figure 3 1 The Calibration Screen ...
Страница 29: ...1 INTRODUCTION Figure 3 2 A Calibration Report ADMINISTRATION 3 4 ...
Страница 31: ...1 INTRODUCTION Figure 3 3 The High Voltage Adjustment Screen ADMINISTRATION 3 6 ...
Страница 32: ...1 INTRODUCTION Figure 3 4 The High Voltage Report 3 7 ADMINISTRATION ...
Страница 34: ...1 INTRODUCTION Figure 3 5 The Chi Square Screen 3 9 ADMINISTRATION ...
Страница 35: ...1 INTRODUCTION Figure 3 6 The Chi Square Report ADMINISTRATION 3 10 ...
Страница 37: ...1 INTRODUCTION Figure 3 7 An Administration Report ADMINISTRATION 3 12 ...
Страница 39: ...ADMINISTRATION 3 14 1 INTRODUCTION Figure 3 8 The Isotope Editing Screen ...
Страница 40: ...1 INTRODUCTION 3 15 ADMINISTRATION Figure 3 9 The Select Isotope Window ...
Страница 42: ...1 INTRODUCTION 3 17 ADMINISTRATION Figure 3 10 The Detector Efficiency Screen ...
Страница 44: ...1 INTRODUCTION Figure 3 11 The Geometric Efficiency Window 3 19 ADMINISTRATION ...
Страница 48: ...1 INTRODUCTION 3 23 ADMINISTRATION Figure 3 13 A Spectrum Analysis Report ...
Страница 50: ...1 INTRODUCTION Figure 4 1 The Procedure Definition Screen THYROID UPTAKE 4 2 ...
Страница 51: ...1 INTRODUCTION 4 3 THYROID UPTAKE Figure 4 2 The Patient Definition Screen ...
Страница 56: ...THYROID UPTAKE 4 8 1 INTRODUCTION Figure 4 3 The Thyroid Uptake Screen ...
Страница 58: ...THYROID UPTAKE 4 10 1 INTRODUCTION Figure 4 4 A Thyroid Uptake Report ...
Страница 59: ...1 INTRODUCTION Figure 4 5 A Time Activity Report 4 11 THYROID UPTAKE ...
Страница 63: ...MANUAL MODE 5 2 1 INTRODUCTION Figure 5 1 The Manual Mode Screen ...
Страница 64: ...1 INTRODUCTION 5 3 MANUAL MODE Figure 5 2 A Manual Mode Report ...
Страница 66: ...1 INTRODUCTION WIPE TEST 6 2 Figure 6 1 The Wipe Counting Screen ...
Страница 67: ...6 3 WIPE TEST CONTENTS Figure 6 2 The Wipe Area Setup Screen ...
Страница 68: ...CONTENTS WIPE TEST 6 4 Figure 6 3 The Wipe Preferences Window ...
Страница 70: ...1 INTRODUCTION Figure 6 4 The Background LLD Screen WIPE TEST 6 6 ...
Страница 72: ...1 INTRODUCTION WIPE TEST 6 8 Figure 6 5 The Wipe Results Report ...
Страница 76: ...1 INTRODUCTION 7 3 BIOASSAY Figure 7 2 The Bioassay Preferences Screen ...
Страница 78: ...7 5 BIOASSAY CONTENTS Figure 7 3 An Individual Bioassay Report ...
Страница 79: ...CONTENTS Figure 7 4 An Employee Bioassay Summary BIOASSAY 7 6 ...
Страница 80: ...1 INTRODUCTION 7 7 BIOASSAY Figure 7 5 An Individual Bioassay History Summary ...
Страница 82: ...1 INTRODUCTION SCHILLING TESTS 8 2 Figure 8 1 Defining a Facility Standard Schilling Test ...
Страница 83: ...CONTENTS 8 3 SCHILLING TESTS Figure 8 2 The Schillings Counting Page ...
Страница 84: ...CONTENTS SCHILLING TESTS 8 4 Figure 8 3 A sample Schillings Mallinckrodt Report ...
Страница 87: ...CONTENTS 8 7 SCHILLING TESTS Figure 8 4 The Diopac Schillings Counting Page ...
Страница 88: ...SCHILLING TESTS 8 8 CONTENTS Figure 8 5 A Sample Diopac Schillings Report ...
Страница 90: ...CONTENTS HEMATOLOGY 9 2 9 1 The IHSA Counting Page ...
Страница 92: ...1 INTRODUCTION HEMATOLOGY 9 4 Figure 9 2 A Sample IHSA I 125 Test Report ...
Страница 93: ...CONTENTS 9 5 HEMATOLOGY Figure 9 3 The CR 51 Counting Page ...
Страница 95: ...CONTENTS 9 7 HEMATOLOGY Figure 9 4 A Sample Cr 51 Volume Test Report ...
Страница 96: ...CONTENTS Figure 9 5 The Cr 51 Red Blood Cell Survival Counting Page HEMATOLOGY 9 8 ...
Страница 98: ...CONTENTS HEMATOLOGY 9 10 Figure 9 6 A sample Cr 51 Survival Test Report ...
Страница 100: ...CONTENTS Figure 9 7 The GFR Counting Page HEMATOLOGY 9 12 ...
Страница 103: ...1 INTRODUCTION 9 15 HEMATOLOGY Figure 9 8 A One Sample Glomerular Filtration Rate Report ...
Страница 104: ...1 INTRODUCTION HEMATOLOGY 9 16 Figure 9 9 A Two Sample Glomerular Filtration Rate Report ...
Страница 106: ...1 INTRODUCTION HEMATOLOGY 9 18 Figure 9 10 The ERPF Counting Page ...
Страница 108: ...1 INTRODUCTION HEMATOLOGY 9 20 Figure 9 10 The ERPF Counting Page ...
Страница 109: ...1 INTRODUCTION 9 21 HEMATOLOGY Figure 9 11 An Effective Renal Plasma Flow Report ...