The disposal of equipment must be in accordance with the respective 
national regulations. 

Wear parts are considered hazardous waste! After being replaced, wear 
parts must be disposed of according to country-specific waste laws. 

4.

 

Specifications 

Dimensions:

 84.5" l x 25.5" w (214.6 x 64.8 cm) with OR accessory rails 

Tabletop:

 53" l x 21" w (134.6 x 53.3 cm); 84" l (213.4 cm) with extension. Tested and complies 

with IPx4. 

Lightweight Extension (for patient transfer only, not for imaging): 26.5" l x 21" w (67.3 x 53.3 
cm) 
Radiolucent Extension: 

31" l x 25.5" w (78.7 x 64.8 cm) with OR accessory rails

 

Radiolucent Area: 

54" l x 17.5" w (137.2 x 44.5 cm)

 

Tabletop Material: 

Three Fiberesin™ sections – each with individual 2" mattress

 

Mattress: 

Individual, seamless, 2" thick (5 cm), one for each section

 

Motions

Height Adjustable:

 34" to 43.5" (86.4 to 110.5 cm) 

 

Tabletop X Motion (head-to-toe float):

 20" (51 cm) 

Tabletop Y Motion (side-to-side float):

 10" (25.4 cm)  

Trendelenburg:

 0° to 20° 

Reverse Trendelenburg:

 0° to 20° 

 

Lateral Roll

: 0° to 20°  

Fowler Positioning:

 Adjusts from 0° to 80° with touch handle lock/unlock 

Controls

Hand Control:

 activates all motions 

Foot Control:

 activates all motions  

Fuses on Control Box: 

F1           250V, AC, 16A, Fast blow / 6,3x32mm 
F2           250V, AC, 16A, Fast blow / 6,3x32mm 
F3           250V, AC, 2A, Fast blow / 6,3x32mm 

Optional: 

X-Ray Film Tray: 

Removable film cassette tray for 14" x 17" film with cassette mask for 10" x 

12" film. Film cassette tray can be removed from either side of the table.

 

X-Ray Grid: 

Removable X-Ray grid, 103 line, 8:1 ratio

 

Accessory Rails: 

Standard OR accessory rails 1.125" x .375" (2.86 x .95 cm) on table (rails also on 

table imaging extension)

 

Patient Restraints: 

Two body straps, two ankle straps

 

Finish: 

Powder coat

 

Patient Capacity: 

In urologic position: 500 lb (227 kg) 
In use with radiolucent extension: 400 lb (181.4) 
Weight tested to four times the patient load rating. 

Shipping Weight: 

903 lb (409.6 kg)

 

Power: 

115 VAC or 230 VAC and Battery. Battery automatically charges when table is plugged in. 

An additional battery can be charged with optional wall-mounted charger.

 

Certifications: 

IEC60601-1:2005 (Third Edition) + CORR1:2006+Corr 2:2007 + A1:2012 (or IEC 

60601-  1:2012 reprint), ANSI/AAMI ES60601-1:2005+A1:2012+C1:2009+A2:2010, CAN/CSA 
C22.2 No: 60601-1:14

IEC 60601-2-46 Edition 3.0, FDA Class II Equipment, CE Class I Non-

measuring. 
CDRH registered and FDA Device Listed

 

Warranty: 

Two years parts and labor 

Содержание 058-800

Страница 1: ...UROLOGY BRACHYTHERAPY C ARM TABLES INSTRUCTIONS FOR USE 058 800 058 805 058 810 058 815 FN 20 324 CLR 11 20...

Страница 2: ...equest or directly from the Biodex website www biodex com tables800 If the desired information is not found contact a local distributor or Biodex directly at supportservices biodex com Thank you Biode...

Страница 3: ...Made Easy 8 2 Operating Instructions 10 The Locking Brakes and Swivel Casters 10 Fowler Positioning 11 Using the Hand Held Controller 11 Using the Foot Controller 12 Disposable Skirt Optional 13 I V P...

Страница 4: ...4 Definition of Symbols The following symbols and their associated definitions are used and implied throughout this manual...

Страница 5: ...2012 or IEC 60601 1 2012 reprint ANSI AAMI ES60601 1 2005 A1 2012 C1 2009 A2 2010 CAN CSA C22 2 No 60601 1 14 CE Class I Non measuring FDA Class II Equipment IEC 60601 2 46 Edition 3 0 Type B Applied...

Страница 6: ...product are not permitted and will void the manufacturer s warranty Unauthorized modification of the product may result in a hazard to the user and or patient Do not modify this equipment without aut...

Страница 7: ...use in a client environment For product specifications refer to the Table of Contents This medical electrical equipment requires special precautions regarding EMC and must be assembled and placed into...

Страница 8: ...ides capability for vascular cardiac and other general C Arm applications Select from knee crutches or the PAL stirrups arm boards or drain systems Contraindications The Urology and Brachytherapy C Ar...

Страница 9: ...and Film Cassette Tray 7 Rail Mounted I V Pole 8 Head End Locking Brakes and Swivel Casters 9 Vertical Doors for adding transfer or radiolucent extension 10 Battery Pack 11 Arm Boards 12 Foot End Ind...

Страница 10: ...ces and across firm rugs The foot end casters are locked from either side of the table while the head end casters are locked together using the Caster Brake Locking Bar To lock the foot end brakes and...

Страница 11: ...owler Back must be set in one of the eight positions provided To lower the Fowler Back depress and hold the Fowler Back button on the underside of the Fowler Back grip and position the top as desired...

Страница 12: ...Y motion Releasing the rocker switch at any time immediately stops the motion NOTE When the table is used in a wet environment it is recommended that the Foot Controller be covered with a plastic bag...

Страница 13: ...xtension is used Stand at the foot end of the table and press the extension release levers located under the middle of each leg extension rail The extension can be slid out from the table Close the ve...

Страница 14: ...re the pole To adjust the height of the I V pole loosen the height adjustment locking knob and raise or lower the top section of the pole to the appropriate height Tighten the locking knob to secure K...

Страница 15: ...ob on the block move the arm board to the desired position and re tighten the knob Drain Bag Holder Optional A flexible drain bag holder is available for this table To install loosen the clamps on the...

Страница 16: ...f desired the pan can be stored in the table by simply pushing in on the drain pan support bars Figure 2 9 The Drain Pan Properly Installed X Ray Grid Optional The optional removable x ray grid is 103...

Страница 17: ...h mechanism To remove the cassette grasp the tray handle and lift it up 1 4 before sliding it toward the side of the table It is not necessary to remove the tray to remove the cassette There are holes...

Страница 18: ...s and dirt accumulation 3 Periodically inspect all welds 4 Periodically check bolts on table tighten if necessary 5 Periodically inspect all strap holders Any holder that feels loose should be removed...

Страница 19: ...rol activates all motions Foot Control activates all motions Fuses on Control Box F1 250V AC 16A Fast blow 6 3x32mm F2 250V AC 16A Fast blow 6 3x32mm F3 250V AC 2A Fast blow 6 3x32mm Optional X Ray Fi...

Страница 20: ...3 Crutch Knee Simplicity pair 058 860 Drain Pan Stainless Steel 058 862 Stirrups PAL with Boot Pad pair 058 863 Arm Board Dura Board with Pad pair 058 865 Arm Board Standard pair 058 866 Cassette Tray...

Отзывы: