Surface Sampling Device 

 
 

GENERAL DESCRIPTION

 

The LIGHTNING MVP ICON Surface Sampling Device is a 
self-contained device for use with the LIGHTNING MVP 
ICON™ or LIGHTNING

MVP

®

 systems for quality and 

HACCP-related monitoring of the environment. The MVP 
ICON and LIGHTNING MVP systems detect food and 
microbiological residues by measuring Adenosine 
triphosphate (ATP) using luciferase bioluminescence. ATP is 
found in most food residues and all bacteria, yeast and mold 
cells.   

Part No. 64003-100 (100 tests) 

KIT COMPONENTS 

One box of 100 surface sampling devices 

IMPORTANT USAGE GUIDELINES

 

Do not insert wet or contaminated sampling devices into the 
instrument. Wipe condensate from exterior of device before 
use with a clean, lint-free cloth. Avoid handling the device 
with wet or contaminated hands. 

Always remove sampling devices from instrument after 
reading. 

Do not store the MVP ICON or MVP instrument on its side 
or back. Avoid storing in humid areas where water leaks are 
possible. 

TEST PROCEDURE 

A. Collecting Sample

 

(a)

 

Twist to remove sampling device from body tube. The 
swab tip must be moist. If fully saturated, any excess 
wetting agent should be blotted off by gently pressing the 
swab tip against the interior of the body tube. Be careful 
not to touch the swab or inside the sampling device with 
fingers. 

(b)

 

Thoroughly swab a standard 10 x 10 cm (4 x 4 in) area of 
interest for a typical flat surface. For irregularly shaped 
surfaces, ensure swabbing technique remains consistent 
each time you swab. Reinsert the swab into the body 
tube using a twisting motion to ensure it is sealed.  

Note:

  After swabbing, the sampling device is stable for one 

hour prior to activation. 

B. Activating Device and Reading Results

 

(a)

 

To activate the sampling device, hold upright and push 
the plunger down until it is fully depressed. The reagents 
will be dispensed through the sampling device washing 
the sample off the swab tip and into the collection tube. 

(b)  Insert the sampling device into the MVP ICON or MVP 

and read the results. When inserting into the MVP 
ICON, ensure the flat sides of the sampling device base 
chamber face left-to-right in the instrument. 

Note:

  Results should be read within two minutes of 

activation. Please refer to the MVP ICON User Guide or 
MVP System Guide for operating instructions. 

INTREPRETING RESULTS

 

(a)

 

When using default thresholds in the MVP ICON or MVP 
systems, readings of 2.5 Zones or less indicate the 
surface is considered clean.  

(b)

 

Readings between 2.6 - 3.0 Zones indicate a warning the 
surface is not adequately clean.  

(c)

 

If the reading is greater than 3.0 Zones, the surface is 
considered dirty and should be re-cleaned and re-tested. 

CONTROLS 

A.

 

Positive Control 

It is recommended LIGHTNING MVP ICON ATP Positive 
Controls (Part No. 64001-25) be run every other day to 
verify integrity of the reagent system.  

(a)

 

A positive control should yield a result from 3.1 to 4.3 
Zones. See ATP Positive Control Directions for Use for 
additional information. 

B.

 

Negative Control 

A negative control is an unused yet activated LIGHTNING 
MVP ICON Surface Sampling Device. 

(a)

 

Negative control results should read between 1.0 to 2.5 
Zones.  

(b)

 

Readings above 2.5 may indicate contamination of the 
MVP ICON or MVP sampling chamber or sampling 
device.  

See ATP Positive Control Directions for Use for additional 
information. Contact BioControl Technical Support at 
800.245.0113 for more information. 

PRECAUTIONS 

The appearance of a yellow band on the swab tip is normal 
and will not affect performance. 

If accidental activation prior to sampling occurs, do not use 
device. 

Hold sampling device upright when activating. 

Read sampling device within two minutes of activation. 

When reading the sampling device, the MVP ICON or MVP 
must be kept upright (no greater than 45° angle). See MVP 
ICON User Guide or MVP System Guide for additional 
information. 

STORAGE

 

Sampling devices should be stored at a temperature between 
2 ºC – 30 ºC (36 

˚

F – 86 ºF). Protect from direct sunlight. 

Unused devices must be resealed in the foil bags. 

 