Chapter 6 Parameter Monitoring
--86--
User Manual of Veterinary Monitor
Sweep
Select [Sweep]: 6.25 mm/s, 12.5 mm/s and 25.0 mm/s. Set the waveform scanning speed.
Pump Switch
Select [Pump Switch]: Off or On.
6.8.5 Zero
The purpose of zero calibration is to eliminate the influence of baseline drift on results to guarantee the accuracy
of the measured results in the process of measuring.
Side-stream
The side-stream CO
2
module automatically calibrates zero when necessary. Users can also manually
zero-calibrates as desired: [User Maintain>>]-[CO
2
Maintenance>>]-[Zero]. Zero calibration requires
disconnecting the animal airway.
Attention
Don
’t perform zero calibration when the temperature is not stable.
Main-stream
The main-stream CO
2
module need zero calibration in the following cases:
The airway adapter is exchanged.
The sensors is re-connected with the module.
The gas readings are found having errors.
The system prompts ‗CO
2
Need Zero‘.
In this case, please check the airway adapter and make sure its adapter window is not blocked by mucus, etc.
Cleaning or adapter replacement is needed when a blockage is found.
The zero calibration steps are as follows:
1.
Connect the sensor with the CO
2
module.
2.
Select CO
2
parameter region and choose [CO
2
Setup]-[Operate Mode], and set [Measure].
3.
When warming-up is done, mount the sensor on a clean and dry airway adapter. The adapter should be
open to the air and all CO
2
sources should be isolated, including breathing machine, animal‘s respiration
and operator‘s respiration.
4.
Select [CO
2
Maintenance>>] in the menu [User Maintain>>]-[Zero], and ‗CO
2
Zero Progress‘ is
prompted on the screen.
5.
The prompting message disappears when the zero calibration is completed.
Attention
The module must be zero-calibrated when the system prompts that CO
2
measure is over the time
allowed. The module should be zero-calibrated regularly after a long use period.
Zero calibration must be performed by specialized technicians.
Содержание PM-900
Страница 2: ......
Страница 12: ...Content VIII User Manual of Veterinary Monitor The Blank Page ...
Страница 14: ...Preface 2 User Manual of Veterinary Monitor The Blank Page ...
Страница 42: ...Chapter 3 Basic Operation 30 User Manual of Veterinary Monitor The Blank Page ...
Страница 132: ...Chapter 12 Battery 120 User Manual of Veterinary Monitor The Blank Page ...
Страница 142: ...Chapter 15 Troubleshooting and Solutions 130 User Manual of Veterinary Monitor The Blank Page ...
Страница 176: ...Append D Factory Default Setup 164 User Manual of Veterinary Monitor The Blank Page ...
Страница 181: ......