Chapter 6 Parameter Monitoring
--76--
User Manual of Veterinary Monitor
please refer to the Accessory Use Manual for more details.
If the packaging of the invasive pressure sensor marked for disinfection is damaged, please stop
using it and contact the provider of the invasive pressure sensor.
If the pressure sensor and conduit are damped, please stop using it and contact the provider of the
invasive pressure sensor.
Attention
Before monitoring, please zero-calibrate first.
Make sure that the pressure sensors are constantly leveled with the heart during monitoring.
Please continuously inject heparin saline to flush conduits to prevent them from being clogged and
maintain the pressure-measuring paths are smooth while connecting the conduits firmly to prevent
them from any displacement or separation that would affect the IBP measure.
6.7.3 Monitoring Steps
1.
Insert the pressure transducer extension cables into the IBP interface on the monitor‘s left panel.
2.
Prepare flushing solution.
3.
Flush the system and deflate the conduits completely. Make sure no bubbles exists in the sensors and valves.
4.
Connect the pressure tube and animal conduit.
5.
Place the sensor on the same level with the heart, approximately at the midaxillary line.
6.
Select correct label.
7.
Zero-calibrate the sensors. Close the sensor‘s valves toward the air and open the valve in connection with the
animal when the zero-calibration is completed.
Figure 6.22 IBP Monitor Connection Diagram
Warning
Please flush the system again with solution if any bubbles are found in the conduit system to avert
any possible incorrect pressure readings caused by such bubble.
The sensors should be leveled with the ears of a sitting animal while measuring his/her intracranial
pressure. Incorrect positions may result in incorrect pressure readings.
Содержание PM-900
Страница 2: ......
Страница 12: ...Content VIII User Manual of Veterinary Monitor The Blank Page ...
Страница 14: ...Preface 2 User Manual of Veterinary Monitor The Blank Page ...
Страница 42: ...Chapter 3 Basic Operation 30 User Manual of Veterinary Monitor The Blank Page ...
Страница 132: ...Chapter 12 Battery 120 User Manual of Veterinary Monitor The Blank Page ...
Страница 142: ...Chapter 15 Troubleshooting and Solutions 130 User Manual of Veterinary Monitor The Blank Page ...
Страница 176: ...Append D Factory Default Setup 164 User Manual of Veterinary Monitor The Blank Page ...
Страница 181: ......