Chapter 7 Alarm
--94--
User Manual of Veterinary Monitor
Attention
In the ISO mode, the sound interval for the low-level alarm must be bigger than or equal to that of
the mid-level alarm and the sound interval for the mid-level alarm must be bigger than or equal to
that of the high-level alarm.
Voice Alarm
Enter [Alarm Config>>] and select [Voice Alarm]: On or Off.
On: when system alarm, in addition to light and sound alarm, as well as voice alarm.
In addition to
lighting alarm and audible alarm, there is voice alarm.
Off: when system alarm, lights and sound alarm.
Only lighting alarm and audible alarm.
Attention
The Monitor only alarms in Chinese and English. And only sounds the important physiological and
technical voice alarm.
7.7 Alarm Pause
Press the button [PAUSE] on the control panel, then all audible alarms can be stopped, the lighting and parameter
higher/lower limits of the alarm parameter stop flashing, alarm text messages are not displayed, and the display of
the remaining time of the Alarm Pause is displayed in the physiology alarm region and the alarm state symbol is
also displayed.
The monitor directly enters the alarm pause state when it starts, and the pause time can be set in [User
Maintain]-[Alarm Config].
The monitor automatically cancels the alarm pause when the alarm pause time has elapsed. When the monitor is
in the ‗Alarm Pause‘ state, you can press the button [PAUSE] to manually cancel the alarm pause.
Attention
Users intentionally disconnecting sensors, probes or a module can press the button [PAUSE] to
make the system enter the
‘Alarm Pause’ state.
7.8 Alarm Silence
Press the button [SILENCE] on the control panel, then all alarms of the monitor currently in process can be
silenced: the audible alarm and the lighting is cleared while the alarm state symbol
is displayed.
When a physiological alarm is silenced, a ‗√‘ is added in the front of the text message, indicting that the alarm is
silenced, but it is displayed on normal background color and the parameter of the alarm and its upper/lower limits
still keep flashing.
When a technical alarm is silenced, no ‗√‘ is added in the front of the text message and it is displayed with the
background color disappearing.
Under the Alarm Silence state, the text alarm message of a silenced alarm is also cleared when it is no longer
exists.
Содержание PM-900
Страница 2: ......
Страница 12: ...Content VIII User Manual of Veterinary Monitor The Blank Page ...
Страница 14: ...Preface 2 User Manual of Veterinary Monitor The Blank Page ...
Страница 42: ...Chapter 3 Basic Operation 30 User Manual of Veterinary Monitor The Blank Page ...
Страница 132: ...Chapter 12 Battery 120 User Manual of Veterinary Monitor The Blank Page ...
Страница 142: ...Chapter 15 Troubleshooting and Solutions 130 User Manual of Veterinary Monitor The Blank Page ...
Страница 176: ...Append D Factory Default Setup 164 User Manual of Veterinary Monitor The Blank Page ...
Страница 181: ......