User Manual

Содержание Nuancer SR-III

Страница 1: ...User Manual ...

Страница 2: ......

Страница 3: ...paration of urine samples 5 4 2 Preparation of strips 5 Chapter 5 Installation 6 5 1 Preparation 6 5 2 Loading the thermal printing paper 6 Chapter 6 Operation Procedure 8 6 1 Turn On Off 8 6 2 Initiate Analysis 8 6 3 Steps 8 6 4 Installation of barcode reader and scanning of medical record No 10 6 5 Test the Instrument Function with Check Strip 10 Chapter 7 Interpretation of Test Result 12 7 1 Pr...

Страница 4: ...ta transfer setting 15 9 4 Display of software version 15 Chapter 10 Warning Message Error Message and Troubleshooting 16 10 1 Warning Message and Troubleshooting 16 10 2 Error message and Troubleshooting 17 Chapter 11 Specifications 18 Chapter 12 Symbols Information 20 Chapter 13 Warranty Statement 21 Exemption of Lability 21 Safety Reliability and Performance 21 Chapter 14 Service Overhaul 22 ...

Страница 5: ... to the instrument and danger If the enclosed power cord is incompatible with local requirement please contact your local distributor Please place the instrument on a level surface In case of malfunction please do not attempt to repair the instrument Contact the service personnel that sell the instrument to your company if you encounter any problem or contact BioCare service personnel Contact info...

Страница 6: ...lowing parameters in the urine leukocyte nitrite urobilinogen protein pH occult blood specific gravity ketone bilirubin and glucose ascorbic acid Rationale NuAncer SR III Urine Chemistry Analyzer contains a photodiode light sensor and a pair of white LED light source The light sensor detects the color change in the reflective light from urine strips radiated by the LED light source The analyzer re...

Страница 7: ...ter 3 Packaging and Content 3 1 Packaging Each Package Contains NuAncer SR III Urine Chemistry Analyzer Strip Holder DC Transformer and Power Cord Check Strip Thermal Printing Paper User Manual Warranty Card ...

Страница 8: ... to operate the machine and establish setting Display screen Data display input and output Printer Prints out test data PS 2 barcode reader connection port Located at the back of the machine to connect to a barcode reader RS 232 data transfer port Located at the back of the machine to transfer data to a personal computer Power switch Located on the side of the machine to turn on off the power Barc...

Страница 9: ... be brought to room temperature before testing 4 1 2 Gently swirl the samples before testing Do not centrifuge 4 1 3 Do not use urine that contains preservatives 4 1 4 Use a disposable and thoroughly cleaned container to collect urine samples If the container is not clean or contains detergent or antimicrobial residues the test result might be inaccurate 4 2 Preparation of strips 4 2 1 For use wit...

Страница 10: ...the power 5 1 3 Turn on the switch to initiate the booting procedure Turn on the power to enter the booting procedure until the machine is calibrated and enters the main screen 5 2 Loading the thermal printing paper Install the thermal printing paper in the following order 5 2 1 Lift up the cover of the printer by pulling the handle ...

Страница 11: ...7 5 2 2 Install the printing paper Pull out some paper and flatten it before closing the cover ...

Страница 12: ...analyzing with medical record No Scan the medical record No with BARCODE READER Press CLEAR to delete erroneous scan Press TEST button if the scan is correct to initiate analysis Refer to the detail test steps below 6 3 Steps 6 3 1 Selection of Strip Press Strip button on the upper left corner to select strip There are 4 options on the screen Select the leftmost one Strip Reader III System Initial...

Страница 13: ... When loading the test strip into instrument slide the strip along the tray until it touches the end of the trough 6 3 5 Quick button for quick initiation of analysis Press the button to perform the test immediately without entering the patient s medical record No 6 3 6 The strip holder will retract into the machine automatically to start analysis 6 3 7 Upon the completion of analysis the strip ho...

Страница 14: ...y 6 5 Test the Instrument Function with Check Strip If you suspect the obtained results are inadequate use the check strip that came with the instrument to see if the instrument is working properly The check strip container contains two gray strips Use either one for the test 6 5 1 Wash the test strip holder See Chapter 8 Maintenance on page14 NOTE Unless the holder is washed before check measurem...

Страница 15: ...Strip in the strip holder 6 5 6 Press Quick button to initiate CHECK STRIP analysis 6 5 7 Upon the completion of analysis the strip holder will eject automatically The machine will print out the analysis report automatically Pass means the machine passed the test Fail means the machine failed the test Strip Type S 10L S 11 S 11VET CHECK Date 2018 04 19 17 38 SEQ CHECK ID Check Strip Test MEMO CHEC...

Страница 16: ...Fail the instrument may be malfunctioning Contact your distributor Chapter 7 Interpretation of Test Result 7 1 Printout test report The test report is as shown on right diagram Press Feed if desired to add 4 blank lines after the results Date 2018 04 19 17 38 SEQ 001 S 11 ID MEMO BLD Ery uL BIL 2 1 mg dL URO 1 mg dL KET 10 mg dL PRO 30 mg dL NIT mg dL GLU mg dL PH 6 SG 1 030 LEU 25 WBC ul AA 10 mg...

Страница 17: ...rom memory drive 7 2 2 Press Print to print the report and then Back to return to the main screen 7 2 3 Press Down button to access the previous data Press Up to access the next data Press Print to print the report 002 ID 9789860158359 Print Up Down Back 001 Strip Type S 10L Strip Feed Set Mem ...

Страница 18: ...es with a cotton swab Do not touch the white bar As the white bar will not be in contact with the sample avoid rinsing it unless necessary 8 2 2 Hang dry or use a piece of soft cloth to wipe clean the strip holder after rinsing If the white bar is dirty gently wipe clean with a piece of soft cloth Avoid scratching the white bar 8 2 3 Re install the holder in the machine or if the holder cannot be ...

Страница 19: ...Back to set date 9 2 Time setting 9 2 1 Press Time to set time Press the button under Time to increase by an hour or a minute Hold the button to automatically enter scrolling mode 9 2 2 Press Back to set time 9 3 Automatic data transfer setting 9 3 1 Press Trans to enter setting Press ON button to turn on transfer and OFF button to turn off transfer 9 3 2 Press Back to confirm 9 4 Display of softw...

Страница 20: ...6 Timer cell low power Possible Cause Strip holder missing Incorrect loading of strip the strip was not placed all the way to the end of the strip holder or no strip Incorrect loading of strip the strip was not placed all the way to the end of the strip holder or no strip Jamming of strip holder Out of battery Troubleshooting Insert strip holder Re test a new strip Re test a new strip Take care no...

Страница 21: ...on 1 Gear malfunction 2 Strip holder malfunction 1 Smudges on white bar 2 Reboot and calibration of LED 1 Back sensor malfunction 2 Strip holder jammed in the back Troubleshooting 1 Reboot 2 If the message persists contact the distributor or BioCare service personnel 1 Reboot 2 If the message persists contact the distributor or BioCare service personnel 1 Reboot 2 If the message persists contact t...

Страница 22: ...cific gravity ketone nitrite glucose occult blood pH ascorbic acid test items vary with the type of strips 24 character 2 line backlight LCD Photodiode light sensor and a pair of white LED Approx 40 60 tests hour Max 120 tests hour varies with the type of tests Approx 60 seconds test Approx 90 seconds test If the test involves leukocytes 16cm 20cm 11 5cm 1 0 Kg AC to DC transformer Input 100 240VA...

Страница 23: ...rification of the safe state of the equipment after repair Please make sure all of cables position and connection well after repair 3 Do not to position SR III to make it difficult to operate the disconnection device 4 Do not repair SR III before disconnect power Back PS 2 Automatic 90X15 mm 24X2 characters Seiko Thermal Printer built in 58 mm line thermal printer Celsius 5o C 40o C Humidity 20 80...

Страница 24: ...ructions for use Temperature limitation Use by Catalogue number Caution consult accompanying documents Handle with care This side up Fragile Keep away from sunlight Keep dry Symbol Referent Manufacturer Symbol Referent CE mark Humidity limitation Atmospheric pressure limitation Authorised representative in the European Community In vitro diagnostic medical device ...

Страница 25: ...ot approved by BioCare or direct indirect or inevitable damages due to maintenance performed by individuals not authorized by BioCare or any transportation or other expenses or responsibilities due to delay This guarantee does not include any erroneous careless use of or malfunctioned instruments does not include any instruments for which the original BioCare serial label or product identification...

Страница 26: ...thorization RMA code Call 886 3 2616678 if outside of Taiwan The RMA code must be shown on the outside of the shipping carton You should also provide the model serial No and brief reason for the return overhaul 2 Calculation of shipment fee The customer is responsible for the shipping costs to and from BioCare including the Customs fees Emergency Contact Information BioCare Corporation Address 4F ...

Страница 27: ......