Biobase LC-4KC Скачать руководство пользователя страница 1

Table-top

 

Low

 

Speed

 

Centrifuge

 

LC-4KC

 / LC-4KII

User

 

Manual

BIOBASE GROUP

Version 2020.11

Содержание LC-4KC

Страница 1: ...Table top Low Speed Centrifuge LC 4KC LC 4KII User Manual BIOBASE GROUP Version 2020 11...

Страница 2: ...RAL 2 2 PECULIARITY 2 3 PRINCIPLE OF OPERATION 2 4 SPECIFICATION 3 5 PARTS IDENTIFICATION 3 6 USING CONDITIONS 3 7 OPERATION PROCEDURE 3 8 MAINTAIN 4 9 GUARANTEE 5 10 COMMON TROUBLE RESOLUTION 5 11 CI...

Страница 3: ...into a angle rotor symmetrically the rotor will spin on axis after you start the machine and the relative centrifuge force RCF comes into being while rotating It will make the composition at differen...

Страница 4: ...system 4 Cover 5 Rotor 6 Nut 7 Motor 8 Toroidal transformer 9 Electric controlling system 10 Centrifugal chamber 11 Rubber foot PS there are two buttons on the right side The green button is power but...

Страница 5: ...up to your setting time machine stop running If the machine is working do not open the cover Turn off power OFF press the safety lock and take out test tubes 8 MAINTAIN 8 1 Please read the manual boo...

Страница 6: ...locker can not lock during run Door lock Check and Replace 4 There is noise after starting on the machine Tube with liquid not equaling sample or rotor nut loosen Make the tube with equaling sample ti...

Страница 7: ...6 11 CIRCUIT DIAGRAM...

Страница 8: ...nan City Shandong Province China Tel 86 531 81219803 01 Fax 86 531 81219804 Inquiry export biobase com Complaints customer_support biobase cc After sales service service_sd biobase cc service_ivd biob...

Страница 9: ...Table type Low Speed Centrifuge BKC TL4 BKC TL4II BKC TL4III User Manual BIOBASE GROUP Version 2020 09...

Страница 10: ...on principles 3 4 Technical parameters 4 5 The rotor parameters 4 6 The operation procedures 4 7 Safety Notice 10 8 Warranty Service Instruction 11 9 Comment Malfunctions and Obviating Methods 11 10 T...

Страница 11: ...by the application of its special damper This machine also has a self balance function Step 2 Set up the rotor mode the rotating speed and the needed time The special PWM controlled speed control sys...

Страница 12: ...rotating radius r the horizontal distance from the position of the test solutions to the axle center and the rotating speed n and here is the calculation formula RCF 1 118 10 6 n2 r g n The rotating s...

Страница 13: ...e supply cord into the outlet and press down the power switch the power switch is on the left bottom of the machine Step 3 After pressing down the button of STOP OPEN put the rotors on the motor shaft...

Страница 14: ...numbers 3 The operation icon 4 The setting of rotor modes 5 The setting of rotating speed 6 The SET button 7 The RCF button 8 The START button 9 The UP button 10 The DOWN button 11 The temperature set...

Страница 15: ...er the first adjustment panel or touch the screen and click the rotating speed area to enter the same panel As it is shown below Press UP or DOWN button to adjust the rotating speed value Attention Th...

Страница 16: ...ng the UP or DOWN button As it is shown below Press SET button to enter next step and the needed hours can be chosen by pressing the UP or DOWN button As it is shown below 3 On the bottom of the first...

Страница 17: ...hird adjustment panel the adjustment panel of acceleration and deceleration or touch the screen and click the acceleration and deceleration area to enter the same panel As it is shown below Press the...

Страница 18: ...eration pressing the STOP OPEN button can also stops the operation of the machine When the rotating speed reaches 0 the buzzer would make a buzzing sound and the cue words would be displayed on the wi...

Страница 19: ...covered 5 The faceplate should not be uncovered in the situation that the motor and the rotors do not stop the operation completely 6 The power supply has its own earth wire so the null line and the e...

Страница 20: ...f the rotors the machine would automatically run with the maximum rated speed of the rotors 6 If the operating temperature is too high and the cavity temperature is over 55 centigrade the machine cann...

Страница 21: ...ing to the dampproof and quakeproof requirement the motor shaft should be painted a thin layer of lubricating grease and preserved in the well ventilated indoor places with the temperature ranging fro...

Страница 22: ...nan City Shandong Province China Tel 86 531 81219803 01 Fax 86 531 81219804 Inquiry export biobase com Complaints customer_support biobase cc After sales service service_sd biobase cc service_ivd biob...

Страница 23: ...Table type Low Speed Centrifuge BKC TL5 BKC TL5II User Manual BIOBASE GROUP Version 2020 09...

Страница 24: ...on principles 3 4 Technical parameters 4 5 The rotor parameters 4 6 The operation procedures 5 7 Safety Notice 12 8 Warranty Service Instruction 12 9 Comment Malfunctions and Obviating Methods 13 10 T...

Страница 25: ...y the application of its special damper This machine also has a self balance function Step 2 Set up the rotor mode the rotating speed and the needed time The special PWM controlled speed control syste...

Страница 26: ...rotating radius r the horizontal distance from the position of the test solutions to the axle center and the rotating speed n and here is the calculation formula RCF 1 118 10 6n2r g n The rotating sp...

Страница 27: ...100ml Time range 0 99h59min Rotate speed precision 50r min Noise 65dB Power AC 220V 50HZ 10A Dimension L W H 600 540 360mm Weight 35kg 5 The rotor parameters Model Product name Maximum capacity Maxim...

Страница 28: ...avoided so as to protectthe flexible support system of the motor Then place the adapter with testtubes on the rotors and rotate the rotors The rotors should be turnedflexibly the adapter should be rot...

Страница 29: ...rature and non refrigerated centrifuge has no such setting function 12 The display area of rotating speed and centrifugal force 13 The shift button of rotating speed and centrifugal force 14 The area...

Страница 30: ...el As it is shown below Press UP or DOWN button to adjust the rotating speed value Attention The slowest rotating speed is 100r and the fastest is the maximum rated speed of the rotor 50r rotating spe...

Страница 31: ...button and press the SET button again to enter the time adjustment area The needed minutes can be chosen by pressing the UP or DOWN button As it is shown below Press the SET button to enter next step...

Страница 32: ...it is shown below Press the UP or DOWN button to choose the rotor number Attention Please take the real objects as the reference and find the corresponding real rotor pictures and rotor number 4 Pres...

Страница 33: ...to preserve the adjusted value and then withdraw from the adjusted state Step7After checking out the settings mentioned above press the START button to start the operation of the machine During the o...

Страница 34: ...completely the centrifugal process would restart by pressing and holding the IMPULS button Step10After the rotors stop the rotation completely press the STOP OPEN button to uncover the faceplate if th...

Страница 35: ...c shock 7 After the whole separation the centrifuge machine should be cleaned up in time the power switch should be turn off and the power supply should be cut off 8 It should be very careful to take...

Страница 36: ...ture is over 55 centigrade the machine cannot be operated After its cooling the machine goes back to work normally Besides the malfunctions mentioned above please contact the technicians of the manufa...

Страница 37: ...ing to the dampproof and quakeproof requirement the motor shaft should be painted a thin layer of lubricating grease and preserved in the well ventilated indoor places with the temperature ranging fro...

Страница 38: ...nan City Shandong Province China Tel 86 531 81219803 01 Fax 86 531 81219804 Inquiry export biobase com Complaints customer_support biobase cc After sales service service_sd biobase cc service_ivd biob...

Страница 39: ...Table type Low Speed Centrifuge BKC TL6 BKC TL6II User Manual BIOBASE GROUP Version 2020 09...

Страница 40: ......

Страница 41: ...otor body or the hanging cup is cracked it is strictly forbidden to use otherwise it will burst Statement Jinan Biobase Biotech Co Ltd hereinafter referred to as Biobase has the final interpretation o...

Страница 42: ...peration principles 3 4 Technical parameters 4 5 The rotor parameter 5 6 The operation procedures 6 7 Safety Notice 13 8 Warranty Service Instruction 13 9 Comment Malfunctions and Obviating Methods 14...

Страница 43: ...The special PWM controlled speed control system assures its high control accuracy and its fast acceleration and deceleration function which provides different acceleration and deceleration time to be...

Страница 44: ...ated by this formula 27 4 logeRmax logeRmin Ts min n2 r2 Type Rmax The rotating radius of the test solutions having the farthest distance with the axle center Rmin The rotating radius of the test solu...

Страница 45: ...tor 12 10ml 6000 5140 16 15ml 5000 4400 24 10ml 4000 2810 4 250ml 4000 3290 4 50ml 5000 4600 4 100ml 5000 4800 Swing micro plate rotor 2 2 96 holes 4000 2360 BKC TL6 II Angle rotor 12 10ml 6000 5140 S...

Страница 46: ...ace the adapter with testtubes on the rotors and rotate the rotors The rotors should be turnedflexibly the adapter should be rotated freely and there should be no block orcollision Step 4 Place the ce...

Страница 47: ...The display area of rotating speed and centrifugal force 13 The shift button of rotating speed and centrifugal force 14 The area of acceleration and deceleration 15 The INPULS button 16 The Enter butt...

Страница 48: ...100r and the fastest is the maximum rated speed of the rotor 50r rotating speed can be adjusted by each press Pressing the UP or DOWN button fixedly can adjust the speed value continuously As it is s...

Страница 49: ...9 Press the SET button to enter next step and the needed hours can be chosen by pressing the UP or DOWN button As it is shown below...

Страница 50: ...e real objects as the reference and find the corresponding real rotor pictures and rotor number 4 Press the SET button to enter the third adjustment panel the adjustment panel of acceleration and dece...

Страница 51: ...below 5 When finishing every step mentioned above press the ENTER button for about 2 seconds to preserve the adjusted value and then withdraw from the adjusted state 7 After checking out the settings...

Страница 52: ...IMPULS button the rotors begin the deceleration until they stop rotating During the deceleration process the IMPULS button cannot be pressed any longer when the rotors stop the rotation completely the...

Страница 53: ...ic shock 7 After the whole separation the centrifuge machine should be cleaned up in time the power switch should be turn off and the power supply should be cut off 8 It should be very careful to take...

Страница 54: ...ature is over 55 centigrade the machine cannot be operated After its cooling the machine goes back to work normally Besides the malfunctions mentioned above please contact the technicians of the manuf...

Страница 55: ...g to the damp proof and quake proof requirement the motor shaft should be painted a thin layer of lubricating grease and preserved in the well ventilated indoor places with the temperature ranging fro...

Страница 56: ...nan City Shandong Province China Tel 86 531 81219803 01 Fax 86 531 81219804 Inquiry export biobase com Complaints customer_support biobase cc After sales service service_sd biobase cc service_ivd biob...

Страница 57: ...Table type High Speed Centrifuge BKC MH16 User Manual BIOBASE GROUP Version 2020 10...

Страница 58: ...n principles 3 IV Technical parameters 4 V The rotor parameter 4 VI The operation procedures 5 VII Safety Notice 11 VIII Warranty Service Instruction 11 IX Comment Malfunctions and Obviating Methods 1...

Страница 59: ...the application of its special damper This machine also has a self balance function Step 2 Set up the rotor mode the rotating speed and the needed time The special PWM controlled speed control system...

Страница 60: ...e rotating radius r the horizontal distance from the position of the test solutions to the axle center and the rotating speed n and here is the calculation formula RCF 1 118 10 6n2r g n The rotating s...

Страница 61: ...Table type high speed centrifuge Angle rotor 12 1 5 2 0ml 16500 Angle rotor 24 1 5 2 0ml 14000 Maximum rotate speed 16500r min Maximum centrifugal force 18780 g Maximum capacity 12 10ml Time range 0...

Страница 62: ...p 4 Place the centrifuge tubes into the test tubes The centrifuge tubes should be placed symmetrically and the test solutions in the tubes should be even so as to make sure the equilibrium operation o...

Страница 63: ...he SET button to enter the first adjustment panel or touch the screen and click the rotating speed area to enter the same panel As it is shown below Press the UP or DOWN button to adjust the rotating...

Страница 64: ...N button and press the SET button again to enter the time adjustment area The needed minutes can be chosen by pressing the UP or DOWN button As it is shown below Press the SET button to enter next ste...

Страница 65: ...it is shown below Press the UP or DOWN button to choose the rotor number Attention Please take the real objects as the reference and find the corresponding real rotor pictures and rotor number 4 Pres...

Страница 66: ...current centrifugal force can be displayed on the window by pressing the RCF button or touching the screen and clicking the RCF icon Repress the button and the rotating speed value can be displayed ag...

Страница 67: ...ld not be removed because of its high rebound force against the seal ring on top of the centrifugal chamber please press the faceplate slightly with one hand and repress the STOP OPEN button with the...

Страница 68: ...c shock 7 After the whole separation the centrifuge machine should be cleaned up in time the power switch should be turn off and the power supply should be cut off 8 It should be very careful to take...

Страница 69: ...ture is over 55 centigrade the machine cannot be operated After its cooling the machine goes back to work normally Besides the malfunctions mentioned above please contact the technicians of the manufa...

Страница 70: ...ng to the dampproof and quakeproof requirement the motor shaft should be painted a thin layer of lubricating grease and preserved in the well ventilated indoor places with the temperature ranging from...

Страница 71: ...nan City Shandong Province China Tel 86 531 81219803 01 Fax 86 531 81219804 Inquiry export biobase com Complaints customer_support biobase cc After sales service service_sd biobase cc service_ivd biob...