Biosafety Cabinet Class II A2
BSC-4FA2-GL/BSC-6FA2-GL
User Manual
BIOBASE GROUP
Version 2021.06
Страница 1: ...Biosafety Cabinet Class II A2 BSC 4FA2 GL BSC 6FA2 GL User Manual BIOBASE GROUP Version 2021 06...
Страница 2: ...d Warranty before operating this unit to assure proper operation After reading these documents be sure to store them securely together with the Warranty at a hand place for future reference Warning Be...
Страница 3: ...concept 14 2 1 2 Application Range 14 2 1 3 Air filtration system 14 2 1 4 User instructions 14 2 1 5 Circuit of digital control 15 2 1 6 Airflow velocity indication 15 2 1 7 Differential pressure ind...
Страница 4: ...fection 23 2 6 Replacement part list 23 2 7 Wiring diagram 25 3 Common faults solution 26 3 1 Common faults solution 26 3 1 1 Warning and reminder 26 3 1 2 Trouble shooting 26 3 1 3 Simple accessories...
Страница 5: ...photos 1 1 Unpacking Choose the proper unpacking method according to the actual situation Method 1 Necessary tools for unpacking Electric drill with hexagon dead M8 Picture 1 Method 2 Use M8 Wrench to...
Страница 6: ...rticle as below 4 Unpack the entire package put away all components for assembly carefully Left adhesive tape can be wiped with alcohol 5 Check all parts If there is damage please notify the transport...
Страница 7: ...tive tube 12 5A Placed in a transparent plastic bag 2pcs Protective tube 6 3A 3pcs Protective tube 2 5A 2pcs Protective tube 125Ma 110V Protective tube 63mA 220V 1pc Remote control including battery 1...
Страница 8: ...Ambient temperature 15 35 3 Relative Humidity 75 4 Atmospheric pressure range 70 kPa 106 kPa 5 Electrical parameters Consistent with the rated voltage of the Biosafety Cabinet See 2 1 11 technical pa...
Страница 9: ...alve and check the drain valve in the closed position in parallel with the ground position as shown in figure 7 11 The top of the device is equipped with HEPA exhaust protective device if equipment re...
Страница 10: ...9 Picture 5 Aligned the mounting holes on the bottom of the cabinet side with the locating bolts slow down the cabinet on the mounting base Picture 6...
Страница 11: ...tening nut 5 Rubber gasket 6 Drain valve 7 Copper joints d Adjustment of Footmaster Caster Picture 8 Clockwise rotate caster s red part to low down the base feet and the height of the cabinet Low down...
Страница 12: ...voltage 1 Connect the Biosafety Cabinet dedicated power line to power outlet on the ground electric current is 9A Biosafety Cabinet must be the only equipment on the circuit to ensure that does not e...
Страница 13: ...lease contact the manufacturer for you to provide local on site testing services company names On site debugging should include at least the following Application of aerosol detection technology check...
Страница 14: ...ion 3 And it is necessary to have a balance test between them 3 Downflow air velocity The equipment should be balanced by it And generally only the adjustment outlined in section 1 7 required However...
Страница 15: ...ed area The air filtration system is the most important system which consists of blower air duct supply filter and exhaust filter The function of air filtration system is transferring filtered air to...
Страница 16: ...ter are too big there will be audio and visual alarm then we need to replace the filter to ensure the safety of operator e After self examination the equipment will be in standby state f The fan run b...
Страница 17: ...de continuous stable gas flow exhausting outdoor The total exhaust air rate of the exhaust hood is no less than 100m3 h of safety cabinet s gas displacement gas displacement of safety cabinet BSC 4FA2...
Страница 18: ...20V 10 110V 10 Rated frequency 50 Hz 60Hz External Size W D H 1383 775 2295mm 1873 775 2295 mm Working Zone Size W D H 1210 600 660mm 1700 600 660 mm Rated current 9A 9A Total Airflow Volume 590m3 h 8...
Страница 19: ...18 2 2 Structural composition Picture 12 1 Air Outlet safety guard 5 Alarm rest plate 2 Control panel 6 UV lamp 3 LCD display 7 IV bar 4 Fluorescent lamp 8 Base stand 2 3 Control panel Picture 13...
Страница 20: ...e BSC either by pressing the buttons on control panel or using the remote control There are totally 8 common button on control panel The power button Master switch to control other function keys To co...
Страница 21: ...ton There are totally 8 common buttons on control panel 2 4 The Usage of Remote Control In order to solve the problem that is inconvenient for the user to operate in close range operation the Biosafet...
Страница 22: ...and then adjust other minutes and hours position data in the same way c After the time is confirmed the corresponding display lamp lights by selecting the function buttons such as UV d Press the POWER...
Страница 23: ...l material maintenance Form 1 list out the clean method for different materials Note Special coat in the surface of equipment may be damaged if often use power with abrasive compound to do cleanness M...
Страница 24: ...the right disinfection method 2 6 Replacement part list BSC 4FA2 GL Replacement parts list No Designation Specification aba01 Lamp holder T8 FL011 W aba02 UV Lamp T6 30W aba03 Fluorescent Lamp LED 18W...
Страница 25: ...22W abb04 UV lamp ballast TBE 232FWc F40 T8 abb05 Upper filter Exhaust filter 1240 470 69 abb06 Down flow filter Supply filter 1700 470 69 abb07 Fan Wheel DD10 7 3 4P Motor 5SDA39RL abb08 Control pane...
Страница 26: ...V WINDOW H H L Ctrl B L U E 2 0 A W G B L A C K 2 0 A W G B L A C K 1 8 A W G W H I T E 2 0 A W G W H I T E 1 8 A W G LCD EIGHT CORE CABLE A N J I A N Window L N Green 14AWG BLACK 14AWG BLACK 18AWG WH...
Страница 27: ...attention 3 1 2 Trouble shooting Please confirm whether the power is connected or not whether the power cord is obvious damaged whether the fuse is good before the fault diagnosis Faults Measures Equi...
Страница 28: ...them 2 When failures are not occur and the operator can t solve please notify our maintenance department immediately For your safety please do not maintain equipment by yourself 3 The maintenance of t...
Страница 29: ...corresponding models of the UV lamp tubes and then rotary screw the bulb and take it off then take the correspondence type of lamp and put it to the lamp holder and and rotary screw in reverse directi...
Страница 30: ...29 Picture20 3 Glass door super elevation warning label Picture 21 4 MET Picture 22 5 Warning note Picture 23 6 DECONTAMINATE note Picture 24...
Страница 31: ...no responsibility for risks caused by improper operation and man made damages during the warranty period d This warranty Bill does not pay the trip fee of repairing replacement and returning to the b...
Страница 32: ...flow velocity Measured by Air Flow Anemometer Model type Inflow Velocity Downflow Velocity The reading of the Highest point and lowest point BSC 4FA2 GL A2 0 505 0 555m s 0 305 0 355m s Average 0 08m...
Страница 33: ...nan City Shandong Province China Tel 86 531 81219803 01 Fax 86 531 81219804 Inquiry export biobase com Complaints customer_support biobase cc After sales service service_sd biobase cc service_ivd biob...