20
3.3.2 Edit
Select the program you want to modify in the "Program Manage" interface, and then click [Edit] to
enter the "Program Edit" interface, as shown in Figure 3-8, where you can see each step of the
program. Select a step to edit the parameters of that step.
Figure 3-8
ID
: The sequence of steps in the program. (Automatic generated)
Step Name
: It is recommended to set the name of the step according to the requirements of the
reagent instructions, which can be set freely.
Hole Position
: The operation position of this step, column 1-6 is optional.
(The positions in column 1.6.7.12.13.18.19.24 have heating function, and the temperature control
range is from room temperature to 100°C, that is, positions 1-8 in the instrument marked by
numbers)
Wait Time
: The standing time of this step is minute, and the upper limit is set to 99.
Blend Time
: The shake time of this step is in minutes, and the upper limit is set to 99.
Blend Mode
: Only vertical mode is available.
Blend Rate
: Low, medium and high are available.
Temp
: You can freely set whether this step needs heating and the heating temperature.
Attract Times
: The times that the magnetic rod attracts the beads which can up to 99 times.
Volume
: This volume corresponds to the volume of the solution in the well of the 96-well deep well
plate. The rod sleeve will adjust the shaking distance in lysis and adsorption steps according to the
volume, and calculate the time for one magnetization at this position.
Preheat
: Preheat to ensure that the preset temperature is reached when performing this step.
Top
: Set the selected step as the first step of the program.
Up
: Run the selected step one step ahead.
Down
: Delay the selected step by one step.
Bottom
: Move the selected step to the last step of the procedure.
Add
: Add a new step below the selected step.
Edit
: Turn the above steps into editable state.
Save
: It is usually grayed out. Click to modify and make changes to the steps to become clickable,
and click it to save the changes.