Using Plot and Histogram Tools
To remove hit detection from a region
1.
Point to a plot or histogram that contains the region that uses hit detection.
The plot and histogram toolbar appears.
2.
Click Assign Hit Regions.
3.
In the Available Hit Region dialog box, clear the checkbox of the region for which you want to
remove hit detection.
4.
Point outside the plot or histogram to save your changes.
The target symbol is removed from the region and the region is no longer used for hit detection.
Adding Diagonal Separators to Plots
A diagonal separating line in a density plot can assist you in determining fluorescence overlap and in
setting PMT voltages for a compensation control. In a typical scenario, the parameter to be
compensated is shown on the x-axis (as in the plots that Everest Software creates automatically for
compensation controls). If a positive population falls below the diagonal line, this indicates that the
control sample is brighter in the intended channel than in the channel shown on the y-axis. Thus, the
required compensation will be less than 100%. If a positive population falls above the diagonal line,
this indicates that the control sample is brighter in a channel other than the intended one and will
need more than 100% compensation.
To add a diagonal separator
1.
Point to the plot to which you want to add diagonal aseparator.
The plot toolbar appears.
2.
Click Add Diagonal Separation.
A diagonal line appears on the plot.
User Guide | 249
Содержание ZE5
Страница 1: ...ZE5 Cell Analyzer and Everest Software User Guide Version 2 0 ...
Страница 2: ......
Страница 3: ...ZE5 Cell Analyzer and Everest Software User Guide Software Version 2 0 ...
Страница 14: ...Table of Contents xii ZE5 Cell Analyzer and Everest Software ...
Страница 50: ...Chapter 2 Hardware Description 48 ZE5 Cell Analyzer and Everest Software ...
Страница 82: ...Chapter 4 Everest Software 80 ZE5 Cell Analyzer and Everest Software ...
Страница 84: ...Chapter 4 Everest Software 82 ZE5 Cell Analyzer and Everest Software ...
Страница 92: ...Chapter 4 Everest Software 90 ZE5 Cell Analyzer and Everest Software ...
Страница 104: ...Chapter 4 Everest Software 102 ZE5 Cell Analyzer and Everest Software ...
Страница 109: ...Managing Users 9 Click OK to save the new user account User Guide 107 ...
Страница 120: ...Chapter 5 Configuring the System 118 ZE5 Cell Analyzer and Everest Software ...
Страница 125: ...Working with Optical Filter Configurations User Guide 123 ...
Страница 126: ...Chapter 5 Configuring the System 124 ZE5 Cell Analyzer and Everest Software ...
Страница 173: ...Setting Up the Run List The plate setup displays a layout of the selected media User Guide 171 ...
Страница 177: ...Setting Up the Run List User Guide 175 ...
Страница 191: ...Setting Up the Run List User Guide 189 ...
Страница 276: ...Chapter 7 Creating Experiments and Workspaces 274 ZE5 Cell Analyzer and Everest Software ...
Страница 295: ...Setting Up the Experiment 6 Set up samples User Guide 293 ...
Страница 333: ...Decontaminating the System User Guide 331 ...
Страница 380: ...Appendix B Example 9 Color Immunophenotyping Experiment 378 ZE5 Cell Analyzer and Everest Software ...
Страница 384: ...Appendix C Optional External DI Water and Waste Ports 382 ZE5 Cell Analyzer and Everest Software ...
Страница 392: ...Appendix F References 390 ZE5 Cell Analyzer and Everest Software ...
Страница 397: ......