Auto Compensation Wizard
The compensation wizard assists with establishing the proper compensation matrix
coefficient for each parameter. Single controls for each parameter and the associated
FCS files must be saved prior to compensation.
1.
Run single controls for each parameter and your multi-stained sample.
2.
Save the acquisitions as FCS files.
3.
Go to
Home
->
Compensation Wizard
4.
Select the parameters for compensation
5.
Click the file icon for each selected parameter
6.
Locate and select the single control FCS file for each parameter
7.
Select whether to use height or area for all parameters
8.
Click the forward blue arrow to open the
Automatic Compensation Calculation
Wizard
window