SEQUENTIAL 

CIRCULATOR

 

MODEL SC-3008 

120 West Commercial Avenue, Moonachie, NJ 07074 

Toll-Free Phone: 800-888-0908 / Phone: 201-939-0716 

  Fax: 201-939-4503 

Web site: www.biocompression.com 

E-mail: [email protected] 

Made in U.S.A 

Содержание SC-3008

Страница 1: ...CULATOR MODEL SC 3008 120 West Commercial Avenue Moonachie NJ 07074 Toll Free Phone 800 888 0908 Phone 201 939 0716 Fax 201 939 4503 Web site www biocompression com E mail biosystems biocompression com Made in U S A ...

Страница 2: ... confident that your pump and garment are backed by the best warranty and customer service in the industry By simply dialing our toll free number for customer service 800 888 0908 warranty repairs or adjustments will be performed in a timely manner with minimal inconvenience to you Should a warranty repair require an extended length of time if available a pump from our loaner inventory will be mad...

Страница 3: ...isconnect the garment from the pump to re lease any residual air and pressure in the garment CAUTION The system is not intended for use during sleep INTENDED USE The Model SC 3008 Sequential Circulator is a manual not software driven sequential pneumatic compression device intended for either primary or adjunctive treatment of primary or secondary lymphedema The device is also intended for additio...

Страница 4: ...rotective end caps which envelope each end of the pump thereby suspending the pump on all four sides within the carton This packaging design prevents damage to the pump which would ordinarily be sustained when the carton is thrown or handled roughly by the carriers NOTE The carton and end caps should be saved for re use each time the pump is either transported or shipped When transporting for adde...

Страница 5: ...he inner knob counter clockwise will enable the pressure adjustment knob to again turn free without resistance CAUTION CARE MUST BE TAKEN NOT TO OVER TIGHTEN THE INNER LOCKING KNOB AS ONLY MINIMAL FORCE IS REQUIRED TO LOCK PRESSURE SETTING IN PLACE NOTE CHANGING THE INNER KNOB TO A 1 2 x 8 32 ALLEN HEAD SET SCREW WILL UPGRADE THIS FEATURE TO TAMPER PROOF 3 PRESSURE GAUGE mmHg 4 AIR SUPPLY PORTS NU...

Страница 6: ...Bar from Auxiliary Air Supply Ports 6 under controls indicators and attach the Latch Connector Bar in the same manner ADJUSTABLE GRADIENT PRESSURE NOTE When a pressure adjustment is made to any individual chamber one must be conscious of the overall gradients across all chambers so as not to create reverse gradients 8 LATCH CONNECTOR BAR Numbered White Circle O ...

Страница 7: ... ELECTRIQUES 120 V c a 60 Hz 0 5 A When servicing use only identical replacement parts Do not remove cover Refer to qualified service personnel Lors des reparations utilisez exclusivement des pieces de re change identiques No retirez pas le couvercle Consultez un tech niciaen qualifie WARNING REPLACE FUSE WHERE MARKED ADVERTISSEMENT REMETTRE LE FUSIILE A L ENDROIT INDIQUE Fuse rated 3 Amps Time De...

Страница 8: ...cap and remove fuse After placing new fuse in cap slot push cap and fuse inward and turn clockwise to lock in place NOTE The outer safety fuse is the only item serviceable by someone other than a Bio Compression Sys tems technician at the factory Bio Compression Systems technicians have been trained specifical ly for the manufacturing and repair of all Bio Compression Systems products NOTE Having ...

Страница 9: ...ols and features of this equipment you are now ready to begin your treatment according to your physician s prescribed course of therapy 1 Make sure that your circulator is plugged into a safe properly secured 110 V AC outlet 2 Place unit on a sturdy table or other type surface close to where you will be sitting The unit has non slip rubberized feet on the bottom however placing paper or other item...

Страница 10: ...tolerate 40mmHG These patients with a filter may need to divide their treatment into twice a day 30 minutes per treatment It is suggested that the provider obtain a Nega tive Doppler study from the physician for their records All compression settings should be discussed with the physician It is ultimately his her responsibility to prescribe the setting and it should be written on the prescription ...

Страница 11: ...y shut itself off Note that alt hough the switch light will go out the switch itself will remain in the On position As a result prior to tak ing your next treatment you must first flip the On Off switch down to the Off position wait at least 10 seconds enabling the pump to again re set itself for another one hour treatment cycle then flip the switch up again to the On position at which time you wi...

Страница 12: ...bers 1 2 graduated down to the lowest in chambers 7 8 3 The main PRESSURE ADJUSTMENT KNOB 2 sets the pressure for sections 1 2 of the sleeve As it is adjusted it will also raise or lower in proportion the remaining sections 4 In the unlikely event you would wish to change the gradient settings calibrated at the factory it is pos sible to do so by using the over ride adjustments 8 on the underside ...

Страница 13: ...ARRANTY INFORMATION You can feel confident that your product is backed by the best warranty in the industry covering any and all malfunctions including parts and labor resulting from component and or manufacturing defects Compression Pump 3 years from date of purchase invoice Sleeves Garments 1 year from date of purchase invoice NOTES _____________________________________________ Serial Number ___...

Страница 14: ...emove soil on surface of garment may require additional washing by hand with a clean towel while submerged Avoid using any abrasive materials such as scrubbing pads or chemicals that could cause damage to the exterior surface of garment 7 Re Submerge garment for 30 minutes with exception of tubing connectors in solution consisting of 1 cup of bleach per 1 gallon of warm tap water again agitating g...

Страница 15: ...13 ...

Страница 16: ...8 Timer Issue 01 24 12 Rev E 08 11 14 Other bio compression systems products Also available Sequential Circulator Model 2004 2004 FC Sequential Circulator Model 3004 3004 fc Sequential Circulator model 2008 The Biocryo system Multi Flo DVT Combo Prophylaxis System BioArterial Plus Arterial Blood Flow Enhancement System Compression therapy garments Including our Custom Garments with the fastest tur...