VDL (E3.1) 06/2020
Page 48/206
3.5.2
Area classification in the surroundings of the chamber
The VDL vacuum drying oven with the exception of the power plug is classified in category 3 in relation to
the environment . It may be installed in areas in which explosive atmospheres may occur on a rare and
temporary basis.
The chamber plug (power plug) is unprotected because it must be disconnected in case of an emergency
to completely separate the chamber from the power supply. therefore the electrical connection must be
established outside a zone.
Spreading of an explosive atmosphere, which may occur on a rare and temporary basis, to unprotected
areas must be reliably prevented by measures of the operator. In particular, this includes a sufficiently
dimensioned one technical ventilation according to country-specific regulations (TRBS 2152 Part 2 for
Germany).
Figure 8: Area classification in the surroundings of the chamber (schematic representation)
Occurrence of an explosive atmosphere:
occasionally: in the chamber interior with tubing
on a rare and temporary basis: in the ambiance of the chamber with the exception of the power plug
never (unprotected areas): Connection location of the power plug
Содержание vdl 115
Страница 192: ...VDL E3 1 06 2020 Page 192 206 28 6 Dimensions 28 6 1 VDL 23 Dimensions in mm ...
Страница 193: ...VDL E3 1 06 2020 Page 193 206 28 6 2 VDL 56 Dimensions in mm ...
Страница 194: ...VDL E3 1 06 2020 Page 194 206 28 6 3 VDL 115 Dimensions in mm ...
Страница 197: ...VDL E3 1 06 2020 Page 197 206 30 Certificates and declarations of conformity 30 1 EU Declaration of Conformity ...
Страница 198: ...VDL E3 1 06 2020 Page 198 206 ...
Страница 199: ...VDL E3 1 06 2020 Page 199 206 ...
Страница 200: ...VDL E3 1 06 2020 Page 200 206 31 Product registration ...
Страница 207: ......