Permissible ambient temperature range for transport: -10 °C /
14°F
to +60 °C /
140°F
.
You can order transport packing and rolling pallets for transportation purposes from BINDER Service.
Freezer transport in upright position is mandatory in order to avoid oil running out of the engine casing and
resulting damages to the cooling system. Max. angle of inclination during transport: 10°.
Wear suitable shoes (safety shoes) during transport.
Following transport, wait at least 8 hours until start-up.
3.3
Storage
Intermediate storage of the chamber is possible in a closed and dry room. Observe the guidelines for
temporary decommissioning (chap. 25.2).
•
Permissible ambient temperature range for storage: -10 °C /
14°F
to +60 °C /
140°F
.
•
Permissible ambient humidity: max. 70% r.h., non-condensing
Secure the chamber against unintentional rolling by locking the front castors.
When after storage in a cold location you transfer the chamber to its warmer installation site, condensa-
tion may form in the inner chamber or on the housing. Before start-up, wait at least one hour until the
freezer has attained ambient temperature and is completely dry. According to the type of transport that
has taken place (chap. 3.2) you may have to wait at least 8 hours until start up.
3.4
Location of installation and ambient conditions
The freezer is designed for setting up inside a building (indoor use). Set up the chamber on a flat, even
surface, free from vibration and in a well-ventilated, dry location. Lock the front castors and align the
chamber using a spirit level. The site of installation must be capable of supporting the chamber’s weight
(see technical data, chap. 27.4).
CAUTION
Danger of overheating.
Damage to the chamber.
∅
Do NOT set up chambers in non-ventilated recesses.
Ensure sufficient ventilation for dispersal of the heat.
CAUTION
Leakage of refrigerant in the event of a chamber defect.
Danger to the environment.
Ensure sufficient ventilation of the installation site.
•
Permissible ambient temperature range for operation: +18 °C /
64.4 °F
to +32 °C /
89.6 °F
. At elevated
ambient temperature values, fluctuations in temperature can occur.
The ambient temperature should not be substantially higher than the indicated ambient tem-
perature of +25 °C /
77 °F
to which the specified technical data relate. For other ambient condi-
tions, deviations from the indicated data are possible.
Prevent the freezer from sucking warm air from other devices.
Содержание ULTRA.Guard UF V 500
Страница 106: ...UF V E3 09 2018 Page 106 116 27 7 Dimensions UF V 500 E3 Dimensions in mm...
Страница 107: ...UF V E3 09 2018 Page 107 116 27 8 Dimensions UF V 700 E3 Dimensions in mm...
Страница 108: ...UF V E3 09 2018 Page 108 116 28 Certificates and declarations of conformity 28 1 EU Declaration of Conformity...
Страница 109: ...UF V E3 09 2018 Page 109 116...
Страница 110: ...UF V E3 09 2018 Page 110 116 29 Product registration...
Страница 117: ......