MKF+ MKFT (E3.2) 07/2017
page 20/115
3.2
Guidelines for safe lifting and transportation
The front castors of the chamber can be blocked by brakes. Please move the chambers with castors only
when empty and on an even surface, otherwise the castors may be damaged. Mount the upholstered
steel L-type profile at the lower door locking. After operation please observe the guidelines for temporarily
decommissioning the chamber (chap. 18.2).
CAUTION
Sliding or tilting of the chamber.
Damage to the chamber.
Risk of injury by lifting heavy loads.
Do NOT lift or transport the chamber using the door handle, the door or at the lower
housing.
Do NOT lift the chamber by hand.
Transport the chamber only in its original packaging.
Secure the chamber with transport straps for transport.
Keep the chamber in upright position.
Place the chamber using technical devices (fork lifter) on the transport pallet. Set the
fork lifter only from the rear in the middle of the chamber. Make sure to place all the
lateral supports of the chamber on the forks.
Transport the chamber with the original transport pallet. Set the fork lifter ONLY to the
pallet. Without the pallet the chamber is in imminent danger of overturning.
You can order transport packing and pallets for transportation purposes from BINDER service.
Permissible ambient temperature range during transport:
•
If the steam humidifying system has NOT been emptied: +3 °C /
37.4 °F
to +60 °C /
140 °F
.
•
After BINDER Service has emptied the steam humidifying system: -10 °C /
14 °F
to +60 °C /
140 °F
.
With temperatures below +3 °C /
37.4°F
, water must be completely removed from the humidifying system.
CAUTION
Transport below +3 °C / 37.4°F with filled steam humidifying system.
Freezing in the steam generator.
Damage to the chamber.
Contact BINDER Service before any transportation below +3 °C /
37.4°F
.
3.3
Storage
Intermediate storage of the chamber is possible in a closed and dry room. Observe the guidelines for
temporary decommissioning (chap. 18.2).
Permissible ambient temperature range during storage:
•
If the steam humidifying system has NOT been emptied: +3 °C /
37.4 °F
to +60 °C /
140 °F
.
•
After BINDER Service has emptied the steam humidifying system: -10 °C /
14 °F
to +60 °C /
140 °F
.
Содержание MKF 115
Страница 97: ...MKF MKFT E3 2 07 2017 page 97 115 20 10 Dimensions MKF 115 MKFT 115...
Страница 98: ...MKF MKFT E3 2 07 2017 page 98 115 20 11 Dimensions MKF 240...
Страница 99: ...MKF MKFT E3 2 07 2017 page 99 115 20 12 Dimensions MKFT 240...
Страница 100: ...MKF MKFT E3 2 07 2017 page 100 115 20 13 Dimensions MKF 720 MKFT 720...
Страница 102: ...MKF MKFT E3 2 07 2017 page 102 115...
Страница 103: ...MKF MKFT E3 2 07 2017 page 103 115...
Страница 104: ...MKF MKFT E3 2 07 2017 page 104 115 21 2 EU Declaration of Conformity for MKFT...
Страница 105: ...MKF MKFT E3 2 07 2017 page 105 115...
Страница 106: ...MKF MKFT E3 2 07 2017 page 106 115...
Страница 108: ...MKF MKFT E3 2 07 2017 page 108 115...
Страница 109: ...MKF MKFT E3 2 07 2017 page 109 115 22 Product registration...