KMF (E5.2) 06/2014
page 44/99
9.
Program operation
The MB1 2-channel program controller permits programming temperature and humidity cycles. It offers
25 program memory positions with 100 program sections each. The total cumulative number of program
sections is limited to 500. It is not possible to link several programs.
For each program section a temperature set-point, a humidity set-point, the fan speed (0% to 100%), and
the switching-state of up to 8 operation lines (non-functional with standard unit) can be entered.
Programming is possible directly by the keypad of the controller or graphically by the software APT-
COM™ 3 DataControlSystem (option, chap. 14.1) specially developed by BINDER.
9.1 Menu-based program entry
Displays showing the initial normal display in Idle Mode
W
X
08:43:55 15.12.13
CONFIG
VIEW->
PGM
TEMP
0.0
26.8
°C
HUMID
0.0
58.6
%rH
HAND
W
X
08:43:55 15.12.13
CONFIG
VIEW->
PGM
FAN SPEED
50
%
HAND
X - TEMP
26.8
°C
X - HUMID
58.6
%rH
Hit button PGM. The window
program selection
appears.
Select a program via the arrow keys and confirm by pressing “ENTER”
The following display serves to select a
subroutine
:
TP-Program 1
Entry of temperature values and fan speed setting
TP-Program 2
Entry of humidity values
TP-Program 3
no function
DEL PGM
08:43:55 15.12.13
Prog select Fr. Abs. 372
Prog 1 PROG 01
Prog 2 PROG 02
Prog 3 PROG 03
Prog 4 PROG 04
Prog 5 PROG 05
Prog 6 PROG 06
Prog 7 PROG 07
Prog 8 PROG 08
Prog 9 PROG 09
Prog10 PROG 10
Prog11 PROG 11
Prog12 PROG 12
Prog13 PROG 13
Prog14 PROG 14
Prog15 PROG 15
Prog16 PROG 16
Prog17 PROG 17