15.1 Communication software APT-COM™ 3 DataControlSystem (option)
The climatic chamber is regularly equipped with an Ethernet interface (8) that can connect the BINDER
communication software APT-COM™ 3 DataControlSystem. The actual temperature and humidity values
are given at adjustable intervals. Programming can be performed graphically via PC. Up to 30 chambers
with RS 422 interface can be cross-linked. The MAC Address is indicated below the Ethernet interface.
For further information, please refer to the operating manual of the BINDER communication software
APT-COM™ 3.
The additional RS422 interface (9) is only used for service purposes. Do NOT connect it to any network.
The interface is labeled accordingly.
15.2 Interface RS 422 (option)
With this option, the chamber is equipped with a serial interface RS 422 instead of the Ethernet Interface,
that can connect the BINDER communication software APT-COM™ 3 DataControlSystem. The actual
temperature and humidity values are given at adjustable intervals.
Pin allocation of the RS 422 interface: Pin 2:
RxD (+)
Pin 3:
TxD (+)
Pin 4:
RxD (-)
Pin 5:
TxD (-)
Pin 7:
Ground
15.3 Data logger kits
BINDER Data Logger Kits offer an independent long-term measuring system for temperature and humidi-
ty, available for different temperature ranges. According to the selected kit, the Data Logger can measure
and record also the ambient temperature and humidity values via a second multi-function sensor.
BINDER Data Loggers are equipped with a keyboard and a large LCD display, alarm functions and a real-
time function. Measurement data are recorded in the Data Logger and can be read out after the meas-
urement via the RS232 interface of the Data Logger. It offers a programmable measuring interval and
permits storing up to 64000 measuring values. Reading out is done with the Data Logger evaluation soft-
ware. You can give out a combined alarm and status protocol directly to a serial printer.
Data Logger Kit TH 70:
Multi-function sensor for chamber temperature and humidity: Temperature range
-40 °C /
-40 °F
up to 70 °C /
158 °F
, humidity range 0% r.H. up to 100% r.H.
Data Logger Kit TH 70/70:
Multi-function sensor for chamber temperature and humidity: Temperature
range -40 °C /
-40 °F
up to 70 °C /
158 °F,
humidity range 0% r.H. up to 100% r.H. Multi-function sensor
for ambient temperature and humidity: Temperature range -40 °C /
-40 °F
up to 70 °C /
158 °F,
humidity
range 0% r.H. up to 100% r.H.
For detailed information on installation and operation of the BINDER Data Logger, please
refer to the mounting instructions Art. No. 7001-0204 and to the original user manual of the
manufacturer, supplied with the data logger.
Содержание KBF 115 (E5.2)
Страница 5: ...KBF KBF P E5 2 02 2012 page 5 103 Product registration ...
Страница 98: ...KBF KBF P E5 2 02 2012 page 98 103 19 7 Dimensions KBF 115 ...
Страница 99: ...KBF KBF P E5 2 02 2012 page 99 103 19 8 Dimensions KBF 240 KBF P 240 ...
Страница 100: ...KBF KBF P E5 2 02 2012 page 100 103 19 9 Dimensions KBF 720 KBF P 720 ...