10. User level
In this menu the following parameters can be se (in brackets the corresponding abbreviated information
given in display 2):
•
Chamber address
(Address)
Setting of controller address (1 to 254) for operation with the APT-COM™ 4 Multi Management
Software.
•
User code
(User-cod)
Modification of the user code setting (factory setting “0001”) for access to the user level and the
program editor.
Keep in mind any modification of the user code. Without a valid user code it is no longer
possible to access these levels.
•
Safety controller mode
(Saf.mode)
The set-point mode can be entered as ...
Limit:
Maximally permitted temperature as absolute value (example: temperature set-point 37 °C /
98.6°F
, limit setting to 39 °C /
102.2°F
)
Offset:
Maximally permitted temperature exceeding of the actual set-point temperature (e.g. 2 °C).
This maximally permitted temperature automatically changes with every change of the temperature
set-point.
•
Safety controller set-point
(Saf.setp)
The temperature set-point setting of the safety controller (over temperature safety device class 3.1) is
shown and can be changed. An absolute value (e.g. 40 °C /
104°F
) in case of setting “limit” or a
relative value (e.g. 2 °C) in case of setting “Offset” can be entered.
Regularly check the setting of the set-point mode and of the over temperature safety
device set-point entered in this mode for the temperature value in Fixed value operation or
for the maximum temperature value of the selected temperature program in Program
operation mode.
•
Decimal point position
(Decimal)
Selection if integer values or one position after the decimal point can be entered. The integer
representation is shown in Display 2 (set-point entry) while the actual value in Display 1 is always
displayed with one decimal point.
•
Audio Alert
(Buzzer)
Inactive:
no audible signal (buzzer) in case of an alarm event.
Active:
in case of an alarm event (see chap. 13.2) an audible signal (buzzer) will sound. It can be reset
by pressing the “
EXIT
” button.
•
Selection of controller menu language
(Language)
German
,
English
, or
French
can be selected.
•
Counter of operating hours
(Oper.hs)
Information about the number of operating hours currently reached or since the last reset. (no setting,
display only).
•
Max. number of operating hours
(Op.limit)
Entry of a limit number of operating hours, i.e., the maximum number of operating hours that can be
run. Maximum setting: 9999. Reaching the limit has no effect.
•
Reset operating hours
(Op.back)
Reset operating hours to zero.
Содержание KB 23
Страница 75: ...KB 23 KB 23 UL E3 1 02 2019 page 75 87 20 7 KB KB UL 23 dimensions...
Страница 77: ...KB 23 KB 23 UL E3 1 02 2019 page 77 87...
Страница 78: ...KB 23 KB 23 UL E3 1 02 2019 page 78 87...
Страница 80: ...KB 23 KB 23 UL E3 1 02 2019 page 80 87...
Страница 81: ...KB 23 KB 23 UL E3 1 02 2019 page 81 87 22 Product registration...