CB (E7) 06/2018
Page 125/174
When operating a CB with silicon access ports, both silicon plugs must tightly close the access ports. If
the plugs are inserted in a not-gastight manner, or if plugs are missing, CO
2
/ O
2
/ N
2
gas (chamber with
O
2
control) may escape into the environment. The CO
2
control and the O
2
control (chamber with O
2
control) only turn off when the chamber door is opened.
WARNING
High concentration of CO
2
(> 4 Vol.-%).
Risk of death by suffocation.
Danger of poisoning.
Tightly close each access port with two plugs during operation.
Note for chambers with O
2
control:
WARNING
High concentration of O
2
(> 21 % O
2
).
Fire and explosion hazard through contact of combustible materials with O
2
.
Risk of burns and other injuries.
Tightly close each access port with two plugs during operation.
18.3 Interior socket 230V (option, available via BINDER INDIVIDUAL Customized
Solutions)
The interior socket is activated and deactivated in the controller via the switching
state of an operation line (chap. 7.3, 9.7.3, 10.6.5). When the interior socket is
activated, the notification “Interior socket“ is shown highlighted in black next to the
information icon in the screen header in Normal display (chap. 11.1.1)
The interior socket (G) is located at the upper left corner on the rear wall of the inner chamber. It is closed
with a waterproof lid. The interior socket is suitable to supply electrical devices inside the chamber with
230V AC voltage.
Figure 22: Position of the interior socket 230 V
(G)
Interior socket
(G)